微波软X射线显微术用于5-10 GHz频率范围内的纳米级磁化动力学

Microwave soft x-ray microscopy for nanoscale magnetization dynamics in the 5-10 GHz frequency range

Stefano Bonetti, Roopali Kukreja, Zhao Chen, Detlef Spoddig, Katharina Ollefs, Christian Schöppner, Ralf Meckenstock, Andreas Ney, Jude Pinto, Richard Houanche,… 展开作者

Stefano Bonetti, Roopali Kukreja, Zhao Chen, Detlef Spoddig, Katharina Ollefs, Christian Schöppner, Ralf Meckenstock, Andreas Ney, Jude Pinto, Richard Houanche, Josef Frisch, Joachim Stöhr, Hermann Dürr, Hendrik Ohldag

摘要:

我们提出了一种扫描透射X射线显微镜设置结合一种新的微波同步方案,以研究高频磁化动力学在同步光源。所需的灵敏度,以检测小的变化,在短的时间尺度和纳米空间尺寸的磁化强度是通过结合开发的激发机制与单光子计数电子被锁定到同步加速器的工作频率。所需的机械稳定性通过显微镜的紧凑设计来实现。我们的仪器能够在5-10 GHz范围内创建动态现象的直接图像,分辨率为35 nm。当与圆偏振X射线一起使用时,上述功能可以结合起来研究微波频率下的磁现象,例如铁磁共振(FMR)和自旋波。我们展示了我们的技术的能力,通过呈现相位分辨图像的6 GHz的纳米级自旋波所产生的自旋扭矩振荡器,以及均匀的铁磁旋进与~0.1度振幅在9 GHz的微米尺寸的钴条。

英文摘要:

We present a scanning transmission x-ray microscopy setup combined with a novel microwave synchronization scheme in order to study high frequency magnetization dynamics at synchrotron light sources. The sensitivity necessary to detect small changes of the magnetization on short time scales and nanometer spatial dimensions is achieved by combination of the developed excitation mechanism with a single photon counting electronics that is locked to the synchrotron operation frequency. The required mechanical stability is achieved by a compact design of the microscope. Our instrument is capable of creating direct images of dynamical phenomena in the 5-10 GHz range, with 35 nm resolution. When used together with circularly polarized x-rays, the above capabilities can be combined to study magnetic phenomena at microwave frequencies, such as ferromagnetic resonance (FMR) and spin waves. We demonstrate the capabilities of our technique by presenting phase resolved images of a 6 GHz nanoscale spin wave generated by a spin torque oscillator, as well as the uniform ferromagnetic precession with ~0.1 deg amplitude at 9 GHz in a micrometer-sized cobalt strip.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑