利用光谱X射线成像进行最佳材料鉴别的蒙特卡罗验证

Monte Carlo validation of optimal material discrimination using spectral x-ray imaging

Syen J. Nik, Rune S. Thing, Richard Watts, Tony Dale, Bryn Currie, Juergen Meyer

摘要:

验证以前的工作,在光谱X射线成像的材料歧视的优化报告。使用基于BEAMnrc软件包的蒙特卡罗模拟,对包含多达三种材料的体模的投影图像进行材料分解。首先通过最小化预期计数和模拟计数之间的z分数将模拟投影数据分解为材料基础图像。对BEAMnrc模拟中由造影剂组成的感兴趣区域内的像素进行统计分析。考虑到散射辐射和现实的扫描几何形状,该算法提供的能量箱边界的理论最优值被证明具有$\pm$ 2 keV的精度为2和3材料的分解。最后,对理论模型预测的信噪比进行了验证。因此,可以使用经验证的模型来估计实现特定成像目标所需的每个像素的计数。

英文摘要:

The validation of a previous work on the optimization of material discrimination in spectral x-ray imaging is reported. Using Monte Carlo simulations based on the BEAMnrc package, material decomposition was performed on the projection images of phantoms containing up to three materials. The simulated projection data was first decomposed into material basis images by minimizing the z-score between expected and simulated counts. Statistical analysis was performed for the pixels within the region-of-interest consisting of contrast material(s) in the BEAMnrc simulations. With the consideration of scattered radiation and a realistic scanning geometry, the theoretical optima of energy bin borders provided by the algorithm were shown to have an accuracy of $\pm$2 keV for the decomposition of 2 and 3 materials. Finally, the signal-to-noise ratio predicted by the theoretical model was also validated. The counts per pixel needed for achieving a specific imaging aim can therefore be estimated using the validated model.