arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 1208.2848physics.atom-phphysics.opticsquant-ph
收藏

光晶格钟黑体辐射漂移的高精度校正

High accuracy correction of blackbody radiation shift in an optical lattice clock

Thomas Middelmann, Stephan Falke, Christian Lisdat, Uwe Sterr

摘要:

我们已经确定了黑体辐射在锶中的$5s^2$ $^1$ S $_0$ -- $5s5p$ $^3$ P $_0$时钟跃迁上引起的频移。以前它的不确定性限制了锶晶格钟的不确定性为$1\times10^{-16}$。现在,与黑体辐射位移校正相关的不确定性转化为室温下的$5\times 10^{-18}$相对频率不确定性。我们的评估是基于两个时钟状态的差分直流极化率的测量和使用此值和其他原子属性的实验数据的动态贡献的建模。

英文摘要:

We have determined the frequency shift that blackbody radiation is inducing on the $5s^2$ $^1$S$_0$ -- $5s5p$ $^3$P$_0$ clock transition in strontium. Previously its uncertainty limited the uncertainty of strontium lattice clocks to $1\times10^{-16}$. Now the uncertainty associated to the black body radiation shift correction translates to $5\times 10^{-18}$ relative frequency uncertainty at room temperature. Our evaluation is based on a measurement of the differential dc-polarizability of the two clock states and on a modeling of the dynamic contribution using this value and experimental data for other atomic properties.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑