动态扫描力显微镜中振幅的标定

Calibration of oscillation amplitude in Dynamic Scanning Force Microscopy

Juan Francisco González Martínez, Inés Nieto Carvajal, Jaime Colchero Paetz

摘要:

介绍了一种动态扫描力显微镜中振幅的精确标定方法。它是实验表明,一个典型的电子用于处理的悬臂梁的动态运动,可以调整到转移的悬臂梁运动的热噪声从其谐振频率到一个低得多的频率内的相应的电子的典型带宽。因此,在这种电子设备的同相(“相位”)和异相(“幅度”)输出中测量的热噪声可以与热能$kT$相关。如果悬臂的力常数是已知的,则可以从这些信号中测量的热功率精确地校准振荡幅度。基于这一概念,提出了两种振幅校准方法。一种是基于在用于处理悬臂的动态运动的电子器件的输出处测量的均方根(RMS)的简单计算,另一种是基于相应频谱的分析和品质因数、谐振频率和信号强度的计算。

英文摘要:

A method to precisely calibrate the oscillation amplitude in Dynamic Scanning Force Microscopy is described. It is experimentally shown that a typical electronics used to process the dynamic motion of the cantilever can be adjusted to transfer the thermal noise of the cantilever motion from its resonance frequency to a much lower frequency within the typical bandwidth of the corresponding electronics. Therefore, the thermal noise measured in the in-phase ("phase") and out of phase ("amplitude") output of such an electronics can be related to the thermal energy $kT$. If the force constant of the cantilever is known then the oscillation amplitude can be precisely calibrated from the thermal power measured in these signals. Based on this concept, two procedures for the calibration of the oscillation amplitude are proposed. One is based on simple calculation of the Root Mean Square (RMS) measured at the outputs of the electronics used to process the dynamic motion of the cantilever, and the other one is based on analysis of the corresponding spectrum and the calculation of the quality factor, the resonance frequency and the signal strength.