主动脉覆膜支架弯曲性能的计算比较
Computational comparison of the bending behavior of aortic stent-grafts
摘要:
腹主动脉瘤腔内修复术后的二次干预很常见,因为可能发生覆膜支架(SG)相关并发症(主要是内漏和SG血栓形成)。并发症与SG灵活性不足有关,尤其是在迂曲动脉中展开器械时。关于SG设计和柔性之间的关系知之甚少。因此,本研究的目的是使用有限元分析(FEA)数值模拟两种生产的SG(Aorfix-Lombard Medical(A)和Zenith-Cook Medical Europe(Z))的弯曲。通过评估支架间距变化和横截面变形研究了全局SG行为。定义了四个标准来比较SG的灵活性:最大管腔缩小率、弯曲所需的扭矩、移植物中的最大膜应变和支架中的最大Von Mises应力。对于大于60 \degree的成角,与Z-SG相比,A-SG的这四个标准值较低。总之,根据FEA,A-SG比Z-SG更灵活。A-SG可以降低曲折主髂动脉瘤的并发症发生率。我们的数值模型可以用来评估进一步制造以及新设计的SG的灵活性。
英文摘要:
Secondary interventions after endovascular repair of abdominal aortic aneurysms are frequent because stent-graft (SG) related complications may occur (mainly endoleak and SG thrombosis). Complications have been related to insufficient SG flexibility, especially when devices are deployed in tortuous arteries. Little is known on the relationship between SG design and flexibility. Therefore, the aim of this study was to simulate numerically the bending of two manufactured SGs (Aorfix-Lombard Medical (A) and Zenith-Cook Medical Europe (Z)) using finite element analysis (FEA). Global SG behavior was studied by assessing stent spacing variation and cross-section deformation. Four criteria were defined to compare flexibility of SGs: maximal luminal reduction rate, torque required for bending, maximal membrane strains in graft and maximal Von Mises stress in stents. For angulation greater than 60\degree, values of these four criteria were lower with A-SG, compared to Z-SG. In conclusion, A-SG was more flexible than Z-SG according to FEA. A-SG may decrease the incidence of complications in the setting of tortuous aorto-iliac aneurysms. Our numerical model could be used to assess flexibility of further manufactured as well as newly designed SGs.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。