Gafchromic薄膜剂量学:校准方法和误差分析

Gafchromic film dosimetry: calibration methodology and error analysis

Wouter Crijns, Frederik Maes, Frank Van den Heuvel

摘要:

目的:将水等效Gafchromic EBT 2薄膜的物理透射率参数与透明绝对校准方案中的输送剂量相关联。该方案应易于理解,易于执行,并应能够预测剩余剂量误差。结论:Gafchromic EBT2薄膜的校准是正确的,具有可访问的强大的校准协议。该协议主要处理电影的一致性问题。所提出的方法允许仅使用T0、T_inf和剂量比例因子将剂量与红色通道透射率相关联。基于局部和全局均匀性,可以预测红色通道剂量误差小于5%。

英文摘要:

Purpose : To relate the physical transmittance parameters of the water equivalent Gafchromic EBT 2 Film with the delivered dose in a transparent absolute calibration protocol. The protocol should be easy to understand, easy to perform, and should be able to predict the residual dose error. Conclussions : The gafchromic EBT2 Films are properly calibrated with an accessible robust calibration protocol. The protocol largely deals with the uniformity problems of the Film. The proposed method allowed to relate the dose with the red channel transmittance using only T0, T_inf, and a dose scaling factor. Based on the local and global uniformity the red channel dose errors could be predicted to be smaller than 5%.