The energy resolution of PWO crystals and their dependence on the incoming energy has been investigated. The crystals used were first prototype crystals of a new generation, characterized by a higher light yield. Two crystal types with different shape and light yield, manufactured by suppliers in China and Russia, were used. An experiments were performed on two cooled arrays of 3x3 crystals. Photon beams of eight different energies, ranging from 64 MeV to 715 MeV, were directed into the center crystals. Photo Multiplier Tubes were used to read out the response of the scintillators. The energy resolution was calculated using three different methods - FWHM, a two gaussian fit and a right side fit. The two Gaussian method delivered the lowest sigma value and also the lowest energy resolution. The obtained energy resolution for the Russian crystals was 1.61%/sqrt(E)+0.73% and similarly 1.58%/sqrt(E)+4.25% for the Chinese crystals.
Feasibility studies of the time-like proton electromagnetic form factor measurements with PANDA at FAIR
讨论了用\PANDA探测器在FAIR上测量类时区质子电磁形状因子的可能性。详细模拟了$\bar p +p $湮灭为轻子对以及最重要的背景通道的信号效率。它表明,反应$\bar p +p \to e^++ e^-$的微分截面的精确测量,可以得到在一个广泛的角度和运动学范围。电和磁质子形状因子的模量的单独确定将是可能的,直到$q^2\simeq 14$(GeV/c)$^2$的动量传递平方值。测量总$\bar p +p\to e^++e^-$横截面,直至达到$q^2\simeq 28$(GeV/c)$^2$。从模拟事件得到的结果进行比较,现有的数据。还研究了双光子交换机制的灵敏度。
Realtime calibration of the A4 electromagnetic lead fluoride calorimeter
在粒子物理和核物理中,足够的能量分辨率是量热法的关键问题。MAMI的A4宇称破坏实验的量热计是一个分段量热计,其中事件的能量通过对相邻通道的信号求和来确定。在这种情况下,各个模块的精确匹配对于获得良好的能量分辨率至关重要。我们研制了由1022个氟化铅(PbF_2)晶体组成的全吸收电磁量能器的校准程序。该过程通过求解线性方程组来重建单模块对事件的贡献,涉及1022 x 1022矩阵的求逆。该系统已在300至1500 MeV的束流能量下显示出其功能,并代表了一种新的快速方法,可使量热计永久保持在良好的校准状态。
Development and characterization of a large area silicon pad array for an electromagnetic calorimeter
我们目前的研究和开发工作的第一个版本的6*6阵列的硅垫探测器,在印度进行,为拟议的前向量热计(FOCAL)的ALICE合作升级计划的一部分,在欧洲核子研究中心。主要的动机是开发一个大面积的硅垫阵列实现高能量物理实验的挑战性要求,如低漏电流,高击穿电压,并评估其性能作为原型硅钨(Si-W)电磁(EM)量能器中的有源层。为了实现这些目标,在班加罗尔的Bharat Electronics Limited,在4英寸高电阻率N型晶片上设计了一个36焊盘的硅传感器,其单个焊盘尺寸为1 cm 2。该传感器已被用于组装原型硅钨量热计,并成功地与高能粒子束测试。本文报道了大面积硅传感器的设计、研制及其在实验室用放射源和高能粒子束上的特性。
Kinematic Fitting of electromagnetic calorimeter data - an improved method
Helmholtz-Institut für Strahlen- und Kernphysik and Cluster of Excellence ”Color meets Flavor”, Universität Bonn, Germany(波恩大学斯特拉斯尔与核物理亥姆霍兹研究所及“颜色遇见味道”卓越集群)
Istituto Nazionale di Fisica Nucleare Sezione di Torino(意大利国家核物理研究所都灵分部);Thomas Jefferson National Accelerator Facility(托马斯杰斐逊国家加速器设施);Istituto Nazionale di Fisica Nucleare, Sezione di Genova e Dipartimento di Fisica dell’Università(意大利国家核物理研究所热那亚分部及大学物理系);Istituto Nazionale di Fisica Nucleare, Sezione di CATANIA(意大利国家核物理研究所卡塔尼亚分部);University of Glasgow(格拉斯哥大学);Institut de Physique Nucléaire, CNRS/IN2P3, Université Paris-Saclay(巴黎萨克雷大学CNRS/IN2P3核物理研究所);Istituto Nazionale di Fisica Nucleare, Sezione di Roma Tor Vergata and Dipartimento di Fisica dell’Università di Roma Tor Vergata(意大利国家核物理研究所罗马托维加塔分部及罗马托维加塔大学物理系)
对粲偶素(cbar c)体系的研究是理解强相互作用的有力工具。在用PANDA研究pbarp湮灭过程中,粲偶素态的质量和宽度(如h_c)将得到很高的测量精度,这取决于对动量p和束流分辨率(dp/p= 10 e(-4)-10e(-5))的精确了解,而不受探测器分辨率的限制。通过对hc的分析,证明了精确确定粲介子谱中特定末态的可行性。pbar p -> pi 0 pi 0 pi 0衰变与信号通道pbar p -> h_c -> eta_c + gamma ->(pi 0 + pi 0 + eta)+ gamma竞争的初步背景分析在控制之下。本文比较了粲偶素hc的三种电磁跃迁衰变模式。