用于医疗应用的交互深度增强型Gamma-Ray成像

Depth of Interaction Enhanced Gamma-Ray Imaging for Medical Applications

Christoph W. Lerche

摘要:

一种新的设计,用于一个廉价的深度的相互作用能力的γ射线探测器,适用于核医学应用,特别是正电子发射断层成像,已经开发。该设计利用了连续晶体中闪烁光分布的宽度与伽马射线相互作用的深度之间的强相关性。为了测量分布宽度,人们开发了一种对常用电荷分配电路的廉价修改,可以模拟和瞬时计算二阶矩,并在本工作中提出。该方法已被测试与一个连续的LSO闪烁体的尺寸为42 × 42 × 10立方毫米,光学耦合到紧凑的大面积位置敏感的光电倍增管H8500从滨松的检测器。在所有的非平凡时刻的平均分辨率被发现是小于5%,但他们的直接使用作为估计的三维光转换的位置被证明是不合适的。采用标准的高维多项式插值方法,从测量的矩中重建γ射线的撞击位置。当使用重建的影响的位置,固有的平均空间分辨率的检测器被发现是1.9毫米的横向分量和3.9毫米的相互作用的深度。直接使用裸矩作为位置估计,发现探测器的固有平均空间分辨率分别为3.4 mm和4.9 mm。所需的检测器改进的成本基本上可以忽略不计。

英文摘要:

A novel design for an inexpensive depth of interaction capable detector for gamma rays, suitable for nuclear medical applications, especially Positron Emission Tomography, has been developed. The design takes advantage of the strong correlation between the width of the scintillation light distribution in continuous crystals and the depth of interaction of the gamma-ray. For measuring the distribution width, an inexpensive modification of the commonly used charge dividing circuits that allows analogue and instantaneous computation of the 2nd moment has been developed and is presented in this work. The method has been tested with a detector made of a continuous LSO-scintillator of dimensions 42x42x10 cubic mm and optically coupled to the compact large area position sensitive photomultiplier H8500 from Hamamatsu. The mean resolution in all non-trivial moments was found to be smaller than 5% but their direct use as estimates for the three-dimensional photoconversion position turned out to be unsuitable. Standard polynomial interpolation in higher dimensions has been adopted to reconstruct the impact positions of the gamma-rays from the measured moments. When using the reconstructed impact positions, the intrinsic mean spatial resolution of the detector was found to be 1.9 mm for the transverse components and 3.9 mm for the depth of interaction. Using directly the bare moments as position estimate, the intrinsic mean spatial resolution of the detector was found to be 3.4 mm and 4.9 mm, respectively. The cost for the required detector improvements are essentially negligible.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑