The frequency comb created by a femtosecond mode-locked laser and a microstructured fiber is used to phase coherently measure the frequencies of both the Hg^+ and Ca optical standards with respect to the SI second as realized at NIST. We find the transition frequencies to be f_Hg=1 064 721 609 899 143(10) Hz and f_Ca=455 986 240 494 158(26) Hz, respectively. In addition to the unprecedented precision demonstrated here, this work is the precursor to all-optical atomic clocks based on the Hg^+ and Ca standards. Furthermore, when combined with previous measurements, we find no time variations of these atomic frequencies within the uncertainties of |(df_Ca/dt)/f_Ca| < 8 x 10^{-14} yr^{-1}, and |(df_Hg/dt)/f_Hg|< 30 x 10^{-14} yr^{-1}.
A caesium fountain clock is operated utilizing a microwave oscillator that derives its frequency stability from a stable laser by means of a fiber-laser femtosecond frequency comb. This oscillator is based on the technology developed for optical clocks and replaces the quartz based microwave oscillator commonly used in fountain clocks. As a result, a significant decrease of the frequency instability of the fountain clock is obtained, reaching 0.74E-14 at 100 s averaging time. We could demonstrate that for a significant range of detected atom numbers the instability is limited by quantum projec…
We demonstrated transferring the stability of one highly stable clock laser operating at 729 nm to another less stable laser operating at 698 nm. The two different wavelengths were bridged using an optical frequency comb. The improved stability of the clock laser at 698 nm enabled us to evaluate the systematic frequency shifts of the Sr optical lattice clock with shorter averaging time. We determined the absolute frequency of the clock transition 1S0 - 3P0 in 87Sr to be 429 228 004 229 873.9 (1.4) Hz referenced to the SI second on the geoid via International Atomic Time (TAI).
We report the synthesis and measurement of an ultra-precise and extremely stable optical frequency in the telecommunications window around 1543 nm. Using a fibre-based femtosecond frequency comb we have phase-stabilised a fibre laser at 194 THz to an optical frequency standard at 344 THz, thus transferring the properties of the optical frequency standard to another spectral region. Relative to the optical frequency standard, the synthesised frequency at 194 THz is determined to within 1 mHz and its fractional frequency instability is measured to be less than 2*10^{-15} at 1 s, reaching 5*10^{-…
We present an optical-electronic approach to generating microwave signals with high spectral purity. By circumventing shot noise and operating near fundamental thermal limits, we demonstrate 10 GHz signals with an absolute timing jitter for a single hybrid oscillator of 420 attoseconds (1Hz - 5 GHz).
We report the absolute frequency measurement of the 5s2 1S0-5s5p 3P0 transition in 87Sr optical lattice clock (Sr1) at National Time Service Center (NTSC). Its systematic frequency shifts are evaluated carefully with a total relative uncertainty of 5.1E10-17. The measured absolute frequency is 429 228 004 229 872.91(18) Hz with a relative uncertainty of 4.13E10-16, with reference to the ensemble of primary and secondary frequency standards published in the Circular T bulletin by BIPM through a global navigation satellite system (GNSS) link.
We report on transferring 1E-16-level fractional frequency stability of a "master laser" operated at 1.5 μm to a "slave laser" operated at 698 nm, using a femtosecond fiber comb as transfer oscillator. With the 698 nm laser, the 1S_0 - 3P_0 clock transition of 87Sr was resolved to a Fourier-limited line width of 1.5 Hz (before: 10 Hz). Potential noise sources contributed by the frequency comb are discussed in detail.
With ultracold $^{87}$Sr confined in a magic wavelength optical lattice, we present the most precise study (2.8 Hz statistical uncertainty) to-date of the $^1S_0$ - $^3P_0$ optical clock transition with a detailed analysis of systematic shifts (20 Hz uncertainty) in the absolute frequency measurement of 429 228 004 229 867 Hz. The high resolution permits an investigation of the optical lattice motional sideband structure. The local oscillator for this optical atomic clock is a stable diode laser with its Hz-level linewidth characterized across the optical spectrum using a femtosecond frequency…
The kinetic energy distribution of D$^+$ ions resulting from the interaction of a femtosecond laser pulse with D$_2$ molecules is calculated based on the rescattering model. From analyzing the molecular dynamics, it is shown that the recollision time between the ionized electron and the D$_2^+$ ion can be read from the D$^+$ kinetic energy peaks to attosecond accuracy. We further suggest that more precise reading of the clock can be achieved by using shorter fs laser pulses (about 15fs).
We report measurements of absolute transition frequencies and hyperfine coupling constants for the 8S_{1/2}, 9S_{1/2}, 7D_{3/2}, and 7D_{5/2} states in ^{133}Cs vapor. The stepwise excitation through either the 6P_{1/2} or 6P_{3/2} intermediate state is performed directly with broadband laser light from a stabilized femtosecond laser optical-frequency comb. The laser beam is split, counter-propagated and focused into a room-temperature Cs vapor cell. The repetition rate of the frequency comb is scanned and we detect the fluorescence on the 7P_{1/2,3/2} -> 6S_{1/2} branches of the decay of the …
We demonstrate the use of a fiber-based femtosecond laser locked onto an ultra-stable optical cavity to generate a low-noise microwave reference signal. Comparison with both a liquid Helium cryogenic sapphire oscillator (CSO) and a Ti:Sapphire-based optical frequency comb system exhibit a stability about $3\times10^{-15}$ between 1 s and 10 s. The microwave signal from the fiber system is used to perform Ramsey spectroscopy in a state-of-the-art Cesium fountain clock. The resulting clock system is compared to the CSO and exhibits a stability of $3.5\times10^{-14}τ^{-1/2}$. Our continuously ope…
We report on the setup and characterization of a 698 nm master-slave diode laser system to probe the 1S0-3P0 clock transition of strontium atoms confined in a one-dimensional optical lattice. A linewidth in the order of around 100 Hz of the laser system has been measured with respect to an ultrastable 657 nm diode laser with 1 Hz linewidth using a femtosecond fiber comb as transfer oscillator. The laser has been used to measure the magnetically induced 1S0-3P0 clock transition in 88Sr where a linewidth of 165 Hz has been observed. The transfer oscillator method provides a virtual beat signal b…
The absolute frequency of the 1S0-3P0 clock transition of 87Sr has been measured to be 429 228 004 229 873.65 (37) Hz using lattice-confined atoms, where the fractional uncertainty of 8.6x10-16 represents one of the most accurate measurements of an atomic transition frequency to date. After a detailed study of systematic effects, which reduced the total systematic uncertainty of the Sr lattice clock to 1.5x10-16, the clock frequency is measured against a hydrogen maser which is simultaneously calibrated to the US primary frequency standard, the NIST Cs fountain clock, NIST-F1. The comparison i…
Optical frequency comparison of the 40Ca+ clock transition ν_{Ca} (2S1/2-2D5/2, 729nm) against the 87Sr optical lattice clock transition ν_{Sr}(1S0-3P0, 698nm) has resulted in a frequency ratio ν_{Ca} / ν_{Sr} = 0.957 631 202 358 049 9(2 3). The rapid nature of optical comparison allowed the statistical uncertainty of frequency ratio ν_{Ca} / ν_{Sr} to reach 1x10-15 in only 1000s and yielded a value consistent with that calculated from separate absolute frequency measurements of ν_{Ca} using the International Atomic Time (TAI) link. The total uncertainty of the frequency ratio using optical co…
We demonstrate subpicosecond-time-scale population transfer between the pair of hyperfine ground states of atomic rubidium using a single laser-pulse. Our scheme utilizes the geometric and dynamic phases induced during Rabi oscillation through the fine-structure excited state in order to construct an $X$ rotation gate for the hyperfine-state qubit system. Experiment performed with a femtosecond laser and cold rubidium atoms, in a magneto-optical trap, shows over 98\% maximal population transfer between the clock states.
We report a stability below $7\times 10{}^{-17}$ of two independent optical lattice clocks operating with bosonic ${}^{88}$Sr isotope. The value (429228066418008.3(1.9)${}_{syst}$(0.9)${}_{stat}$~Hz) of the absolute frequency of the ${}^{1}S_{0}$ - ${}^{3}P_{0}$ transition was measured with an optical frequency comb referenced to the local representation of the UTC by the 330 km-long stabilized fibre optical link. The result was verified by series of measurements on two independent optical lattice clocks and agrees with recommendation of Bureau International des Poids et Mesures.
We report on the first absolute transition frequency measurement at the 10^{-15} level with a single, laser-cooled 40Ca+ ion in a linear Paul trap. For this measurement, a frequency comb is referenced to the transportable Cs atomic fountain clock of LNE-SYRTE and is used to measure the S1/2-D5/2 electric-quadrupole transition frequency. After the correction of systematic shifts, the clock transition frequency f_Ca+ = 411 042 129 776 393.2 (1.0) Hz is obtained, which corresponds to a fractional uncertainty within a factor of three of the Cs standard. Future improvements are expected to lead to …
We measured the absolute frequency of the $^1S_0$ - $^3P_0$ transition of $^{171}$Yb atoms confined in a one-dimensional optical lattice relative to the SI second. The determined frequency was 518 295 836 590 863.38(57) Hz. The uncertainty was reduced by a factor of 14 compared with our previously reported value in 2013 due to the significant improvements in decreasing the systematic uncertainties. This result is expected to contribute to the determination of a new recommended value for the secondary representations of the second.
We demonstrate a one-dimensional optical lattice clock with ultracold 171Yb atoms, which is free from the linear Zeeman effect. The absolute frequency of the 1S0(F = 1/2) - 3P0(F = 1/2) clock transition in 171Yb is determined to be 518 295 836 590 864(28) Hz with respect to the SI second.
We stabilise a microwave oscillator at 9.6 GHz to an optical clock laser at 344 THz by using a fibre-based femtosecond laser frequency comb as a transfer oscillator. With a second frequency comb we measure independently the instability of the microwave source with respect to another optical clock laser frequency at 456 THz. The total fractional frequency instability of this optic-to-microwave and microwave-to-optic conversion resulted in an Allan deviation sigma_y, of sigma_y=1.2E-14 at 1 s averaging time (band width 50 kHz). The residual phase noise density is -97 dBc/Hz at 10 Hz offset from …
We present a measurement of the absolute frequency of the 5s$^2$ $^1$S$_0$ to 5s5p $^3$P$_0$ transition in $^{87}$Sr, which is a secondary representation of the SI second. We describe the optical lattice clock apparatus used for the measurement, and we focus in detail on how its systematic frequency shifts are evaluated with a total fractional uncertainty of $1 \times 10^{-17}$. Traceability to the International System of Units is provided via comparison to International Atomic Time (TAI). Gathering data over 5- and 15-day periods, with the lattice clock operating on average 74$\%$ of the time…
We report the direct excitation of the highly forbidden (6s^2) 1S0 - (6s6p) 3P0 optical transition in two odd isotopes of ytterbium. As the excitation laser frequency is scanned, absorption is detected by monitoring the depletion from an atomic cloud at ~70 uK in a magneto-optical trap. The measured frequency in 171Yb (F=1/2) is 518,295,836,593.2 +/- 4.4 kHz. The measured frequency in 173Yb (F=5/2) is 518,294,576,850.0 +/- 4.4 kHz. Measurements are made with a femtosecond-laser frequency comb calibrated by the NIST cesium fountain clock and represent nearly a million-fold reduction in uncertai…
The pursuit of ever more precise measures of time and frequency is likely to lead to the eventual redefinition of the second in terms of an optical atomic transition. To ensure continuity with the current definition, based on a microwave transition between hyperfine levels in ground-state $^{133}$Cs, it is necessary to measure the absolute frequency of candidate standards, which is done by comparing against a primary cesium reference. A key verification of this process can be achieved by performing a loop closure$-$comparing frequency ratios derived from absolute frequency measurements against…
We performed an absolute frequency measurement of the $^1S_{0}$-$^3P_{0}$ transition in $^{87}$Sr with a fractional uncertainty of $1.2 \times 10^{-15}$, which is less than one third that of our previous measurement. A caesium fountain atomic clock was used as a transfer oscillator to reduce the uncertainty of the link between a strontium optical lattice clock and the SI second. The absolute value of the transition frequency is 429 228 004 229 873.56(49) Hz.
A phase-stabilized femtosecond laser comb is directly used for high-resolution spectroscopy and absolute optical frequency measurements of one- and two-photon transitions in laser-cooled \rb atoms. Absolute atomic transition frequencies, such as the 5S$_{1/2}$ F=2 \ra 7S$_{1/2}$ F"=2 two-photon resonance measured at 788 794 768 921(44) kHz, are determined without \textit{a priori} knowledge about their values. Detailed dynamics of population transfer driven by a sequence of pulses are uncovered and taken into account for the measurement of the 5P states via resonantly enhanced two-photon trans…
The highly forbidden $^2$S$_{1/2} \rightarrow ^2$F$_{7/2}$ electric octupole transition in $^{171}$Yb$^+$ is a potential candidate for a redefinition of the SI second. We present a measurement of the absolute frequency of this optical transition, performed using a frequency link to International Atomic Time to provide traceability to the SI second. The $^{171}$Yb$^+$ optical frequency standard was operated for 76% of a 25-day period, with the absolute frequency measured to be 642 121 496 772 645.14(26) Hz. The fractional uncertainty of $4.0 \times 10 ^{-16}$ is comparable to that of the best p…
$^{88}$Sr$^{+}$ optical clock with $7.9\times 10^{-19}$ systematic uncertainty and measurement of its absolute frequency with $9.8\times 10^{-17}$ uncertainty
We have measured the absolute frequency of the optical lattice clock based on $^{87}$Sr at PTB with an uncertainty of $3.9\times 10^{-16}$ using two caesium fountain clocks. This is close to the accuracy of today's best realizations of the SI second. The absolute frequency of the 5s$^2$ $^1$S$_0$-5s5p $^3$P$_0$ transition in $^{87}$Sr is 429,228,004,229,873.13(17) Hz. Our result is in excellent agreement with recent measurements performed in different laboratories worldwide. We improved the total systematic uncertainty of our Sr frequency standard by a factor of five and reach $3\times 10^{-17…
Intense X-ray free-electron lasers (XFELs) can rapidly excite matter, leaving it in inherently unstable states that decay on femtosecond timescales. As the relaxation occurs primarily via Auger emission, excited state observations are constrained by Auger decay. In situ measurement of this process is therefore crucial, yet it has thus far remained elusive at XFELs due to inherent timing and phase jitter, which can be orders of magnitude larger than the timescale of Auger decay. Here, we develop a new approach termed self-referenced attosecond streaking, based upon simultaneous measurements of …
Absolute frequency measurement of the $^{40}$Ca$^{+}$ clock transition using a GNSS link to the SI second
Institut für Quantenoptik und Quanteninformation, Österreichische Akademie der Wissenschaften(奥地利科学院量子光学与量子信息研究所);Universität Innsbruck, Institut für Experimentalphysik(因斯布鲁克大学实验物理研究所);Alpine Quantum Technologies GmbH(阿尔卑斯量子技术有限公司);Singapore University of Technology and Design(新加坡科技设计大学);Qatar Center for Quantum Computing, College of Science and Engineering, Hamad Bin Khalifa University(哈马德·本·哈利法大学科学与工程学院卡塔尔量子计算中心)
The low-lying isomeric state of $^{229}$Th provides unique opportunities for high-resolution laser spectroscopy of the atomic nucleus. We determine the energy of this isomeric state by taking the absolute energy difference between the excitation energy required to populate the 29.2-keV state from the ground-state and the energy emitted in its decay to the isomeric excited state. A transition-edge sensor microcalorimeter was used to measure the absolute energy of the 29.2-keV $γ$-ray. Together with the cross-band transition energy (29.2 keV$\to$ground) and the branching ratio of the 29.2-keV st…
We present the first characterization of the spectral properties of superradiant light emitted from the ultra-narrow, 1 mHz linewidth optical clock transition in an ensemble of cold $^{87}$Sr atoms. Such a light source has been proposed as a next-generation active atomic frequency reference, with the potential to enable high-precision optical frequency references to be used outside laboratory environments. By comparing the frequency of our superradiant source to that of a state-of-the-art cavity-stabilized laser and optical lattice clock, we observe a fractional Allan deviation of $6.7(1)\time…
The frequency ratio of the 1S0(F=1/2)-3P0(F=1/2) clock transition in 171Yb and the 1S0(F=9/2)-3P0(F=9/2) clock transition in 87Sr is measured by an optical-optical direct frequency link between two optical lattice clocks. We determined the ratio (ν_{Yb}/ν_{Sr}) to be 1.207 507 039 343 340 4(18) with a fractional uncertainty of 1.5x10^{-15}. The measurement uncertainty of the frequency ratio is smaller than that obtained from absolute frequency measurements using the International Atomic Time (TAI) link. The measured ratio agrees well with that derived from the absolute frequency measurement re…
The absolute frequency of the $^{87}{\rm Sr}$ lattice clock transition was evaluated with an uncertainty of $1.1\times 10^{-15}$ using a frequency link to the international atomic time (TAI). The frequency uncertainty of a hydrogen maser used as a transfer oscillator was reduced by homogeneously distributed intermittent measurement over a five-day grid of TAI. Three sets of four or five days measurements as well as systematic uncertainty of the clock at $8.6\times 10^{-17}$ have resulted in an absolute frequency of $^{87}{\rm Sr}\ {}^1S_0 - {}^3P_0$ clock transition to be 429 228 004 229 872.8…
The optical lattice clock NICT-Sr1 regularly reports calibration measurements of the international timescale TAI. By comparing measurement results to the reports of eight Primary Frequency Standards, we find the absolute frequency of the $^{87}\mathrm{Sr}$ clock transition to be f(Sr)=$429\,228\,004\,229\,873.082(76)$, with a fractional uncertainty of less than 1.8x10$^{-16}$ approaching the systematic limits of the best realization of SI second. Our result is consistent with other recent measurements and further supported by the loop closure over the absolute frequencies of $^{87}\mathrm{Sr}$…
We describe an optical frequency stabilization scheme of a microwave oscillator that is used for the interrogation of primary caesium fountain clocks. Because of its superior phase noise properties, this scheme, which is based on an ultrastable laser and a femtosecond laser frequency comb, overcomes the frequency instability limitations of fountain clocks given by the previously utilized quartz-oscillator-based frequency synthesis. The presented scheme combines the transfer of the short-term frequency instability of an optical cavity and the long-term frequency instability of a hydrogen maser …
We experimentally investigate an optical clock based on $^{171}$Yb ($I=1/2$) atoms confined in an optical lattice. We have evaluated all known frequency shifts to the clock transition, including a density-dependent collision shift, with a fractional uncertainty of $3.4 \times 10^{-16}$, limited principally by uncertainty in the blackbody radiation Stark shift. We measured the absolute clock transition frequency relative to the NIST-F1 Cs fountain clock and find the frequency to be 518 295 836 590 865.2(0.7) Hz.
Optical atomic clocks are poised to redefine the SI second, thanks to stability and accuracy more than one hundred times better than the current microwave atomic clock standard. However, the best optical clocks have not seen their performance transferred to the electronic domain, where radar, navigation, communications, and fundamental research rely on less stable microwave sources. By comparing two independent optical-to-electronic signal generators, we demonstrate a 10 GHz microwave signal with phase that exactly tracks that of the optical clock phase from which it is derived, yielding an ab…
Global Positioning System (GPS) dissemination of frequency standards is ubiquitous at present, providing the most widespread time and frequency reference for the majority of industrial and research applications worldwide. On the other hand, the ultimate limits of the GPS presently curb further advances in high-precision, scientific and industrial applications relying on this dissemination scheme. Here, we demonstrate that these limits can be reliably overcome even in laboratories without a local atomic clock by replacing the GPS with a 642-km-long optical fiber link to a remote primary caesium…
We have performed for the first time direct laser spectroscopy of the 1S0-3P0 optical clock transition at 265.6 nm in fermionic isotopes of neutral mercury laser-cooled in a magneto-optical trap. Spectroscopy is performed by measuring the depletion of the magneto-optical trap induced by the excitation of the long-lived 3P0 state by a probe at 265.6 nm. Measurements resolve the Doppler-free recoil doublet allowing for a determination of the transition frequency to an uncer- tainty well below the Doppler-broadened linewidth. We have performed absolute measurement of the frequency with respect to…
We apply direct frequency-comb spectroscopy, in combination with precision cw spectroscopy, to measure the ${\rm 4s4p} ^3P_1 \to {\rm 4s5s} ^3S_1$ transition frequency in cold calcium atoms. A 657 nm ultrastable cw laser was used to excite atoms on the narrow ($γ\sim 400$ Hz) ${\rm 4s^2} ^1S_0 \to {\rm 4s4p} ^3P_1$ clock transition, and the direct output of the frequency comb was used to excite those atoms from the ${\rm 4s4p} ^3P_1$ state to the ${\rm 4s5s} ^3S_1$ state. The resonance of this second stage was detected by observing a decrease in population of the ground state as a result of at…
We report absolute frequency measurements of 81 hyperfine components of the rovibrational transitions of molecular iodine at 578 nm using the second harmonic generation of an 1156-nm external-cavity diode laser and a fiber-based optical frequency comb. The relative uncertainties of the measured absolute frequencies are typically $1.4\times10^{-11}$. Accurate hyperfine constants of four rovibrational transitions are obtained by fitting the measured hyperfine splittings to a four-term effective Hamiltonian including the electric quadrupole, spin-rotation, tensor spin-spin, and scalar spin-spin i…
We report on the development and phase noise performance of a 9.1926 GHz microwave frequency synthesizer to be used as the local oscillator for a Cs fountain clock. It is based on frequency multiplication and synthesis from an ultralow phase noise 5 MHz Oven Controlled Crystal Oscillator (OCXO) and 100 MHz Voltage Controlled Crystal Oscillator (VCXO).The key component of the frequency multiplication is a non-linear transmission-line (NLTL) used as a frequency comb generator. The phase noise of the synthesizer is improved by carefully optimizing the input power, the input and output impedances …
The absolute frequencies of the hyperfine components of the R(26) 62-0 transition in molecular iodine at 501.7 nm are measured for the first time with an optical clockwork based on a femtosecond laser frequency comb generator. The set-up is composed of an Ar+ laser locked to a hyperfine component of the R(26) 62-0 transition detected in a continuously pumped low-pressure cell (0.33 Pa). The detected resonances show a linewidth of 45 kHz (half-width at half-maximum). The uncertainty of the frequency measurement is estimated to be 250 Hz.
The absolute frequency of the $^{87}{\rm Sr}$ clock transition measured in 2015 was reevaluated using an improved frequency link to the SI second. The scale interval of International Atomic Time (TAI) that we used as the reference was calibrated for an evaluation interval of five days instead of the conventional interval of one month which is regularly employed in Circular T. The calibration on a five-day basis removed the uncertainty in assimilating the TAI scale of the five-day mean to that of the one-month mean. The reevaluation resulted in the total uncertainty of $10^{-16}$ level for the …
An Al$^+$ clock with $1.6\times10^{-18}$ systematic uncertainty and its frequency ratios
Physikalisch-Technische Bundesanstalt(德国联邦物理技术研究院);Institut für Quantenoptik, Leibniz Universität Hannover(汉诺威莱布尼茨大学量子光学研究所);Laboratoire Kastler Brossel, Sorbonne Université, CNRS, ENS-PSL Research University, Collège de France(索邦大学、法国国家科学研究中心、巴黎文理研究大学、法兰西公学院卡斯特尔-布罗塞尔实验室)
We demonstrate a one-dimensional optical lattice clock with a spin-polarized fermionic isotope designed to realize a collision-shift-free atomic clock with neutral atom ensembles. To reduce systematic uncertainties, we developed both Zeeman shift and vector light-shift cancellation techniques. By introducing both an H-maser and a Global Positioning System (GPS) carrier phase link, the absolute frequency of the $^1S_0(F=9/2) - {}^3P_0(F=9/2)$ clock transition of the $^{87}$Sr optical lattice clock is determined as 429,228,004,229,875(4) Hz, where the uncertainty is mainly limited by that of the…
We perform spectroscopic observations of the 698-nm clock transition in $^{87}$Sr confined in an optical lattice using a laser linewidth transfer technique. A narrow-linewidth laser interrogating the clock transition is prepared by transferring the linewidth of a master laser (1064 nm) to that of a slave laser (698 nm) with a high-speed controllable fiber-based frequency comb. The Fourier-limited spectrum is observed for an 80-ms interrogating pulse. We determine that the absolute frequency of the 5s$^{2}$ $^{1}$S$_{0}$ - 5s5p $^{3}$P$_{0}$ clock transition in $^{87}$Sr is 429 228 004 229 872.…
We demonstrated for the first time the characterization of absolute frequency stability of three reference cavities by cross beating three laser beams which are independently locked to these reference cavities. This method shows the individual feature of each reference cavity, while conventional beatnote measurement between two cavities can only provide an upper bound. This method allows for numerous applications such as optimizing the performance of the reference cavity for optical clockwork.
We report the operation of a dual Rb/Cs atomic fountain clock. 133Cs and 87Rb atoms are cooled, launched, and detected simultaneously in LNE-SYRTE's FO2 double fountain. The dual clock operation occurs with no degradation of either the stability or the accuracy. We describe the key features for achieving such a simultaneous operation. We also report on the results of the first Rb/Cs frequency measurement campaign performed with FO2 in this dual atom clock configuration, including a new determination of the absolute 87Rb hyperfine frequency.
Aided by ultra-high resolution spectroscopy, the overall systematic uncertainty of the $^{1}S_{0}$-$^{3}P_{0}$ clock resonance for lattice-confined $^{87}$Sr has been characterized to $9\times10^{-16}$. This uncertainty is at a level similar to the Cs-fountain primary standard, while the potential stability for the lattice clocks exceeds that of Cs. The absolute frequency of the clock transition has been measured to be 429,228,004,229,874.0(1.1) Hz, where the $2.5\times10^{-15}$ fractional uncertainty represents the most accurate measurement of a neutral-atom-based optical transition frequency…
Tests of local position invariance (LPI) made by comparing the relative redshift of atomic clocks based on different atoms have been carried out for a variety of pairs of atomic species. In most cases, several absolute frequency measurements per year are used to look for an annual signal, resulting in tests that can span on order of a decade. By using the output of continuously running clocks, we carry out LPI tests with comparable or higher precision after less than 1.5 years. These include new measurements of the difference in redshift anomalies β for hyperfine transitions in Rb87 and Cs133 …