使用实时三维CMR-MOTUS对有心律失常患者进行连续心室容积量化

Feasibility of Continuous Ventricular Volumetric Quantification in Arrhythmias using Real-Time 3D CMR-MOTUS

University Medical Center Utrecht(乌得勒支大学医学中心);The Ohio State University(俄亥俄州立大学)

Thomas E. Olausson, Maarten L. Terpstra, Rizwan Ahmad, Edwin Versteeg, Casper Beijst, Yuchi Han, Marco Guglielmo, Birgitta K. Velthuis, Cornelis van den Berg, Alessandro Sbrizzi

AI总结:

本文提出了一种基于实时三维运动场重建的方法,用于对有心律失常患者进行连续心室容积量化,揭示了传统方法无法检测到的功能异质性。

摘要:

传统心血管磁共振(CMR)动态成像依赖于将多个心跳合并成一个心脏周期,这在心律失常患者中失效,因为心跳间的变化导致运动伪影和功能信息的丢失。实时二维成像捕捉单个心跳,但缺乏对映射心律失常心脏动态的容积覆盖。我们提出了一种三维实时运动场重建方法,通过自由运行的CMR协议,使在有早搏(PVCs)患者中能够进行连续的容积评估。CMR-MOTUS被扩展为联合重建实时三维运动场和一个运动校正的参考图像,从连续、无门控、非呼吸保持的数据中获取,使用变量密度的Cartesian OPRA轨迹。通过将单个分割传播到所有帧,利用重建的运动场计算心跳间射血分数(EF)。该方法在心脏运动仿真实验中进行了验证,并在四名健康志愿者和四名PVC患者中进行了测试。仿真实验EF值与真实值非常接近(22.1±0.6百分比 vs. 21.9百分比)。在健康志愿者中,EF值与二维参考一致,且分布狭窄,反映了生理一致性。在PVC患者中,EF分布呈双峰,低峰对应于具有明显降低EF的PVC心跳。心电图证实了EF不规则性和PVC发作之间的对齐。这些结果表明,三维实时运动场重建能够连续地在心律失常中进行心跳间容积量化,揭示出传统分组所掩盖的功能异质性。双峰EF分布捕捉了PVC的真实血流动力学影响,可能为监测和治疗评估提供临床相关指标。

英文摘要:

Conventional cardiovascular magnetic resonance (CMR) cine imaging combines data across multiple heartbeats, an assumption that can fail in arrhythmia because beat-to-beat variation causes motion artifacts and obscures functional heterogeneity. Although 2D real-time cine resolves individual beats, stacked slices are suboptimal for capturing complex 3D cardiac dynamics. We investigated continuous beat-to-beat volumetric quantification in patients with premature ventricular contractions (PVCs) using free-running 3D real-time CMR. CMR-MOTUS was extended to jointly reconstruct time-resolved 3D motion fields and a motion-corrected reference image from continuously acquired data without breath-holding or ECG gating. Data were acquired with a variable-density Cartesian trajectory and either 3D spoiled gradient-echo or balanced steady-state free-precession imaging. Ventricular volumes were obtained by propagating one manual reference-image segmentation through all reconstructed frames. The method was evaluated in a cardiac motion phantom with static ground-truth acquisitions, 10 healthy volunteers, and 10 patients with PVCs; all in vivo scans were performed after contrast administration. Phantom ejection fraction (EF) agreed closely with ground truth (17.86% versus 17.27%). Healthy volunteers showed narrow beat-to-beat EF distributions, whereas patients with PVCs showed broader and sometimes bimodal distributions. Simultaneous ECG recordings supported the temporal correspondence between volume irregularities and PVC episodes. Free-running joint 3D motion-field and image reconstruction enables continuous beat-to-beat volumetric assessment and can reveal functional heterogeneity obscured by gated or heartbeat-averaged methods. Larger studies are required to establish clinical validity and determine its role alongside standard 2D cine analysis.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑