用于 O4 观测运行的 Virgo 牛顿校准系统

The Virgo Newtonian calibration system for the O4 observing run

Florian Aubin, Eddy Dangelser, Dimitri Estevez, Alain Masserot, Benoît Mours, Thierry Pradier, Antoine Syx, Pierre Van Hove

AI总结:

本文介绍为 Virgo O4 观测运行安装并持续运行的 NCal 绝对校准系统,其注入信号幅度的 1σ 不确定度达 0.17%,优于 PCal,并分析了各项不确定度来源。

摘要:

在先前观测运行期间进行初步测试之后,研究人员开发了一套牛顿校准器(NCal)系统,并为 O4 观测运行安装在 Virgo 引力波探测器上。该系统持续运行,为 Virgo 在本次运行期间提供绝对校准。其注入信号幅度的 1σ 不确定度为 0.17%,优于光子校准器(PCal)等其他校准技术所获得的结果。本文介绍了该 NCal 系统,并详细说明了不同的不确定度来源。

英文摘要:

After initial tests performed during previous observing runs, a Newtonian Calibrator (NCal) system was developed and installed on the Virgo gravitational wave detector for the O4 observing run. This system, which is continuously operated, provides the absolute calibration of Virgo for this run. Its $1\mbox{-}σ$ uncertainty of $0.17\%$ on the amplitudes of the injected signals is better than that obtained with other calibration techniques like the photon calibrator (PCal). This paper presents this NCal system and details the different sources of uncertainties