磁弛豫成像测量的贝叶斯设计

Bayesian design of measurements for magnetorelaxometry imaging

Tapio Helin, Nuutti Hyvönen, Jarno Maaninen, Juha-Pekka Puska

AI总结:

本文针对磁弛豫成像中激活线圈位置选择问题,提出基于贝叶斯最优实验设计的顺序优化方法,结合高斯与全变差先验并利用滞后扩散迭代近似后验,实验表明非对称目标下优化位置可显著提升重建精度。

摘要:

磁弛豫成像的目的是通过测量磁性纳米粒子产生的叠加磁场的弛豫来确定其在受试者体内的分布,此前这些纳米粒子先被外部激活磁场对齐,随后该激活磁场被关闭。本工作将贝叶斯最优实验设计技术应用于(顺序地)选择激活线圈的位置,以在简化的测量设置中提高数据价值并实现更精确的重建。对纳米粒子的分布同时考虑了高斯先验模型和全变差先验模型。前者允许对多个连续激活同时进行离线优化设计计算,而后者通过利用先前测量的数据来选择下一次激活的位置,从而为算法引入了自适应性。全变差先验具有理想的边缘增强特性,但其缺点是失去了计算上具有吸引力的后验密度高斯形式。为克服这一挑战,采用滞后扩散迭代来提供近似的高斯后验模型,并使得标准贝叶斯A-最优性和D-最优性准则也能用于全变差先验。在若干样本目标上进行了二维数值实验,结论是当目标分布或感兴趣区域在形状上非对称时,优化后的激活位置通常比对称参考设置能带来更好的重建结果。

英文摘要:

The aim of magnetorelaxometry imaging is to determine the distribution of magnetic nanoparticles inside a subject by measuring the relaxation of the superposition magnetic field generated by the nanoparticles after they have first been aligned using an external activation magnetic field that has subsequently been switched off. This work applies techniques of Bayesian optimal experimental design to (sequentially) selecting the positions for the activation coil in order to increase the value of data and enable more accurate reconstructions in a simplified measurement setup. Both Gaussian and total variation prior models are considered for the distribution of the nanoparticles. The former allows simultaneous offline computation of optimized designs for multiple consecutive activations, while the latter introduces adaptability into the algorithm by using previously measured data in choosing the position of the next activation. The total variation prior has a desirable edge-enhancing characteristic, but with the downside that the computationally attractive Gaussian form of the posterior density is lost. To overcome this challenge, the lagged diffusivity iteration is used to provide an approximate Gaussian posterior model and allow the use of the standard Bayesian A- and D-optimality criteria for the total variation prior as well. Two-dimensional numerical experiments are performed on a few sample targets, with the conclusion that the optimized activation positions lead, in general, to better reconstructions than symmetric reference setups when the target distribution or region of interest are nonsymmetric in shape.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑