arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 2302.04939physics.med-phcs.NAmath.NAq-bio.TO
收藏

瓣叶材料特性对反流二尖瓣模拟功能的影响

The effects of leaflet material properties on the simulated function of regurgitant mitral valves

Wensi Wu, Stephen Ching, Patricia Sabin, Devin W. Laurence, Steve A. Maas, Andras Lasso, Jeffrey A. Weiss, Matthew A. Jolley

AI总结:

针对无创获取患者特异性瓣叶材料特性困难的问题,研究通过多组正常与反流二尖瓣模型的有限元模拟,开发全自动反流口量化方法,发现材料特性软度变化不超15%时瓣膜指标相对排序稳定,为无精确组织特性下的房室瓣功能定性比较提供依据。

摘要:

三维成像技术的进步使得构建和分析有限元(FE)模型以评估房室瓣的生物力学行为与功能成为可能。然而,尽管如今已能获取患者特异性的瓣膜几何形态,但通过无创方式测量患者特异性瓣叶材料特性几乎仍无法实现。瓣膜几何形态与组织特性均对瓣膜动力学调控起着重要作用,由此引出核心问题:在不精确掌握组织特性的情况下,能否通过房室瓣的有限元分析获得具有临床相关性的见解。为此,我们研究了1)组织可扩展性的影响,以及2)本构模型参数与瓣叶厚度对模拟瓣膜功能和力学性能的作用。我们对比了1个正常二尖瓣(MV)模型和3个具有常见反流机制(瓣环扩张、瓣叶脱垂、瓣叶牵拉)的反流二尖瓣模型的功能指标(如瓣叶对合情况、反流口面积)与力学指标(如应力、应变),涵盖中度与重度两种反流程度。我们开发了一种全新的全自动方法,可精准量化复杂瓣膜几何形态的反流口面积。研究发现,当使用的材料特性比代表性成人二尖瓣本构模型软至多15%时,一组瓣膜的力学与功能指标的相对排序保持不变。我们的研究结果表明,即便在材料特性尚不明确的人群中,有限元模拟也可用于定性比较瓣膜结构的差异与改变如何影响房室瓣的相对功能。

英文摘要:

Advances in three-dimensional imaging provide the ability to construct and analyze finite element (FE) models to evaluate the biomechanical behavior and function of atrioventricular valves. However, while obtaining patient-specific valve geometry is now possible, non-invasive measurement of patient-specific leaflet material properties remains nearly impossible. Both valve geometry and tissue properties play a significant role in governing valve dynamics, leading to the central question of whether clinically relevant insights can be attained from FE analysis of atrioventricular valves without precise knowledge of tissue properties. As such we investigated 1) the influence of tissue extensibility and 2) the effects of constitutive model parameters and leaflet thickness on simulated valve function and mechanics. We compared metrics of valve function (e.g., leaflet coaptation and regurgitant orifice area) and mechanics (e.g., stress and strain) across one normal and three regurgitant mitral valve (MV) models with common mechanisms of regurgitation (annular dilation, leaflet prolapse, leaflet tethering) of both moderate and severe degree. We developed a novel fully-automated approach to accurately quantify regurgitant orifice areas of complex valve geometries. We found that the relative ordering of the mechanical and functional metrics was maintained across a group of valves using material properties up to 15% softer than the representative adult mitral constitutive model. Our findings suggest that FE simulations can be used to qualitatively compare how differences and alterations in valve structure affect relative atrioventricular valve function even in populations where material properties are not precisely known.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑