机器人超声扫描与手动扫描脊柱侧凸评估的可靠性比较

Reliability of Robotic Ultrasound Scanning for Scoliosis Assessment in Comparison with Manual Scanning

The Hong Kong Polytechnic University(香港理工大学)

Maria Victorova, Heidi Hin Ting Lau, Timothy Tin-Yan Lee, David Navarro-Alarcon, Yongping Zheng

摘要:

背景:超声(US)成像用于脊柱侧凸评估对于无经验的操作者具有挑战性。机器人扫描的开发是为了通过深度学习来遵循脊柱弯曲,并向患者的背部施加一致的力。方法:对23例脊柱侧凸患者进行超声扫描,包括机器人扫描和手动扫描。两名人类评分员在机器人和手动冠状图像上测量每个受试者的棘突角(SPA)。结果:机器人方法显示出较高的内部(ICC > 0.85)和评级者间(ICC > 0.77)的可靠性。与手动方法相比,机器人方法在测量冠状面畸形角度时无显著差异(p < 0.05)。对于测量角度总数的最小值86%和最大值97%,评定者内分析的MAD在0 °至5 °的可接受范围内。结论:本研究表明,在机器人扫描获得的超声图像上测量的脊柱侧凸畸形角度与手动扫描获得的角度相当。

英文摘要:

Background: Ultrasound (US) imaging for scoliosis assessment is challenging for a non-experienced operator. The robotic scanning was developed to follow a spinal curvature with deep learning and apply consistent forces to the patient' back. Methods: 23 scoliosis patients were scanned with US devices both, robotically and manually. Two human raters measured each subject's spinous process angles (SPA) on robotic and manual coronal images. Results: The robotic method showed high intra- (ICC > 0.85) and inter-rater (ICC > 0.77) reliabilities. Compared with the manual method, the robotic approach showed no significant difference (p < 0.05) when measuring coronal deformity angles. The MAD for intra-rater analysis lies within an acceptable range from 0 deg to 5 deg for a minimum of 86% and a maximum 97% of a total number of the measured angles. Conclusions: This study demonstrated that scoliosis deformity angles measured on ultrasound images obtained with robotic scanning are comparable to those obtained by manual scanning.