arXiv 2401.02539cs.ROcs.CV
收藏

基于虚拟夹具的机器人辅助深静脉血栓超声检查

Robot-Assisted Deep Venous Thrombosis Ultrasound Examination using Virtual Fixture

Chair for Computer Aided Medical Procedures and Augmented Reality (CAMP), Technical University of Munich (TUM)(慕尼黑工业大学计算机辅助医疗程序与增强现实主席(CAMP));Bristol Robotics Laboratory, University of the West of England(西英格兰大学布里斯托尔机器人实验室);College of Biosystems of Engineering and Food Science, Zhejiang University(浙江大学生物系统工程与食品科学学院);Institute of Biological and Medical Imaging, Helmholtz Zentrum Munchen(慕尼黑亥姆霍兹中心生物医学影像研究所);Department for Vascular and Endovascular Surgery, rechts der Isar University Hospital, Technical University of Munich(慕尼黑工业大学伊斯河畔大学医院血管与血管外科)

Dianye Huang, Chenguang Yang, Mingchuan Zhou, Angelos Karlas, Nassir Navab, Zhongliang Jiang

AI总结:

本文提出一种基于虚拟夹具的机器人超声系统,通过混合力-运动控制和路径优化,实现DVT按压检查的自动化,减少操作者依赖性,并在仿生手臂上验证了系统性能。

摘要:

深静脉血栓(DVT)是一种常见的血管疾病,深静脉内形成血凝块,可能阻塞血流,甚至引发危及生命的肺栓塞。典型的DVT超声(US)检查方法是通过按压目标静脉直至其管腔完全闭合。然而,这种按压检查高度依赖操作者的经验。为减少操作者内和操作者间的差异,我们提出了一种机器人超声系统,采用新颖的混合力-运动控制方案,确保位置和力的跟踪精度,以及探头在目标表面的软着陆。此外,提出了一种基于路径的虚拟夹具,以实现病变位置重复按压操作时的人机交互简便性。为确保不同检查中获得的生物测量结果具有可比性,我们利用外部RGBD相机和超声图像,以从粗到精的方式确定6D扫描路径。首先使用RGBD相机在物体上提取粗略扫描路径。然后,利用超声图像中分割出的血管管腔来优化扫描路径,以确保目标物体的可见性。为生成用于开发虚拟夹具的连续扫描路径,提出了一种考虑位置和方向的基于弧长的路径拟合模型。最后,在具有不平整表面的人体仿生手臂模型上对整体系统进行了评估。

英文摘要:

Deep Venous Thrombosis (DVT) is a common vascular disease with blood clots inside deep veins, which may block blood flow or even cause a life-threatening pulmonary embolism. A typical exam for DVT using ultrasound (US) imaging is by pressing the target vein until its lumen is fully compressed. However, the compression exam is highly operator-dependent. To alleviate intra- and inter-variations, we present a robotic US system with a novel hybrid force motion control scheme ensuring position and force tracking accuracy, and soft landing of the probe onto the target surface. In addition, a path-based virtual fixture is proposed to realize easy human-robot interaction for repeat compression operation at the lesion location. To ensure the biometric measurements obtained in different examinations are comparable, the 6D scanning path is determined in a coarse-to-fine manner using both an external RGBD camera and US images. The RGBD camera is first used to extract a rough scanning path on the object. Then, the segmented vascular lumen from US images are used to optimize the scanning path to ensure the visibility of the target object. To generate a continuous scan path for developing virtual fixtures, an arc-length based path fitting model considering both position and orientation is proposed. Finally, the whole system is evaluated on a human-like arm phantom with an uneven surface.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑