用于安全机器人超声成像的柔顺准直接驱动末端执行器的设计与评估

Design and Evaluation of a Compliant Quasi Direct Drive End-effector for Safe Robotic Ultrasound Imaging

Duke University(杜克大学);University of Wisconsin-Madison(威斯康星大学麦迪逊分校)

Danyi Chen, Ravi Prakash, Vincent Y. Wang, Zacharias Chen, Sarah Dias, Daniel M. Buckland, Leila J. Bridgeman, Siobhan R. Oca, Brian P. Mann

AI总结:

本文设计了一种采用准直接驱动执行器的新型柔顺末端执行器,经离体实验验证,其力跟踪误差及多项图像质量指标均优于传统方案,可提升机器人超声成像的安全性与可靠性。

摘要:

机器人辅助超声扫描有望推进自主且可及的医学成像,但确保患者安全及探头接触时的柔顺人机交互是重大挑战。现有系统多机械刚度高,或在柔顺性上牺牲性能。本文提出一种新型柔顺末端执行器,用于安装在机械臂上实现安全精准的机器人超声成像,采用准直接驱动(quasi-direct drive)执行器实现被动机械柔顺性与精确主动力控制。为评估其性能,开发了离体动态运动模拟器平台,用于模拟运动下组织的接触与扫描测试。以UR3e机械臂采用传统力控制策略为基准开展评估:2.5N至15N的单点接触实验显示,该末端执行器平均降低力跟踪均方根误差80.1%;不同速度下的轨迹扫描实验显示,平均降低力跟踪误差68.0%;在6项定量图像质量与稳定性指标中,有4项观察到统计学显著的改进。本工作提出了一种设计与评估柔顺末端执行器的新方法,旨在提升机器人超声成像的安全性与可靠性。

英文摘要:

Robot-assisted ultrasound scanning promises to advance autonomous and accessible medical imaging. However, ensuring patient safety and compliant human-robot interaction during probe contact poses a significant challenge. Most existing systems either have high mechanical stiffness or trade performance for compliance. This paper presents a novel compliant end-effector designed to mount on robotic arms for safe and accurate robotic ultrasound imaging, using a quasi-direct drive actuator to achieve passive mechanical compliance and precise active force control. To evaluate the end-effector's performance, we developed an ex vivo dynamic motion simulator platform for contact and scanning testing on tissue under simulated movements. The end-effector was evaluated against a UR3e robot arm using conventional force control strategies as a baseline. Single-point contact experiments from 2.5 N to 15 N show that the end-effector reduced force tracking RMS error by 80.1% on average. Trajectory scanning experiments at different speeds showed an average of 68.0% reduction in force tracking error. Statistically significant improvements were observed in four of six quantitative image quality and stability metrics using the end-effector. This work presents a novel approach for designing and evaluating compliant end-effectors, with the goal of improving safety and reliability in robotic ultrasound.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑