偏轴 compliant RCM 连接器:近各向同性刚度与最小寄生误差

Off-Axis Compliant RCM Joint with Near-Isotropic Stiffness and Minimal Parasitic Error

Istituto Italiano di Tecnologia(意大利技术研究院);Politecnico di Milano(米兰理工大学);Università di Genova(热那亚大学);Delft University of Technology(代尔夫特理工大学)

Federico Mariano, Elena De Momi, Giovanni Berselli, Jovana Jovanova, Just L. Herder, Leonardo S. Mattos

AI总结:

本文提出一种偏轴monolithic compliant RCM关节,结合近各向同性刚度与最小寄生运动,用于神经内窥镜操作。通过设计两阶段流程,实现各向同性刚度与RCM漂移的平衡,并通过有限元分析验证。

摘要:

本文提出了一种偏轴、整体compliant的远程运动中心(RCM)关节,用于神经内窥镜操作,结合近各向同性刚度与最小寄生运动。基于Tetra II概念,末端执行器置于四面体柔性结构外部,以提高视线、便于灭菌和允许快速工具释放。设计分为两个阶段:首先通过基于顺应性的各向同性目标确定运动面板尺寸,然后通过有限元可行性探索合成约束面板,以在刚度各向同性和RCM漂移之间进行权衡。该关节通过梁元素建模,并通过详细的有限元分析验证,包括疲劳受限的应力约束。通过选择性激光熔融技术制造了PA12原型,并在台架上进行表征:在末端执行器上施加2 N径向负载,同时6自由度电磁传感器记录姿态。所选配置产生刚度椭圆主轴比(PAR)为1.37,寄生到有用旋转比(PRR)为0.63%。在4.5°命令旋转下,预测的RCM漂移保持亚毫米级(0.015-0.172 mm)。疲劳分析预测了可使用的旋转工作空间为12.1°-34.4°,取决于方向。实验再现了模拟的各向异性刚度趋势,典型偏差为6-30%,证明了紧凑、可制造的RCM模块,适用于受限的手术访问。

英文摘要:

This paper presents an off-axis, monolithic compliant Remote Center of Motion (RCM) joint for neuroendoscopic manipulation, combining near-isotropic stiffness with minimal parasitic motion. Based on the Tetra II concept, the end-effector is placed outside the tetrahedral flexure to improve line of sight, facilitate sterilization, and allow rapid tool release. Design proceeds in two stages: mobility panels are sized with a compliance-based isotropy objective, then constraining panels are synthesized through finite-element feasibility exploration to trade stiffness isotropy against RCM drift. The joint is modeled with beam elements and validated via detailed finite-element analyses, including fatigue-bounded stress constraints. A PA12 prototype is fabricated by selective laser sintering and characterized on a benchtop: a 2 N radial load is applied at the end-effector while a 6-DOF electromagnetic sensor records pose. The selected configuration produces a stiffness-ellipse principal axis ratio (PAR) of 1.37 and a parasitic-to-useful rotation ratio (PRR) of 0.63%. Under a 4.5° commanded rotation, the predicted RCM drift remains sub-millimetric (0.015-0.172 mm). Fatigue analysis predicts a usable rotational workspace of 12.1°-34.4° depending on direction. Experiments reproduce the simulated directional stiffness trend with typical deviations of 6-30%, demonstrating a compact, fabrication-ready RCM module for constrained surgical access.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑