实时水中氚监测仪的研制

Development of a real-time tritium-in-water monitor

C. D. R. Azevedo, A. Baeza, E. Chauveau, J. A. Corbacho, J. Díaz, J. Domange, C. Marquet, M. Martínez-Roig, F. Piquemal, D. Prado, J. F. C. A. Veloso, N. Yahlali

AI总结:

本文研制了一种模块化实时水中氚监测仪,目标达到欧盟饮用水标准100 Bq/L的MDA,单模块实测外推MDA为3.6 kBq/L,已可用于核电站环境实时预警。

摘要:

本文报告了一种用于实时水中氚活度监测的探测器模块的研制与性能。该监测仪基于模块化探测单元,其数量可根据所需灵敏度进行选择。整套系统旨在实现水中氚活度100 Bq/L的最低可探测活度(MDA),该限值由欧盟理事会指令2013/51/Euratom针对人类饮用水所规定。同一系统还可用作核电站水中氚环境监测的实时预警系统。首个探测器模块在完成表征、调试后,立即安装在塔霍河上Arrocampo大坝(西班牙Almaraz核电站)的排放渠之后。由于单个探测模块具有高灵敏度,系统需要通过主动和被动屏蔽技术来抑制放射性本底。我们外推出单个模块的MDA为3.6 kBq/L,该值受宇宙射线本底限制。单个模块所获得的数值已与实时环境监测和预警系统的要求兼容。进一步优化单模块灵敏度将意味着减少模块数量并降低探测器系统的成本。

英文摘要:

In this paper, we report the development and performance of a detector module envisaging a tritium-in-water real-time activity monitor. The monitor is based on modular detection units whose number can be chosen according to the required sensitivity. The full system is being designed to achieve a \emph{Minimum Detectable Activity} (\emph{MDA}) of 100\,Bq/L of tritium-in-water activity which is the limit established by the E.U. Council Directive 2013/51/Euratom for water intended for human consumption. The same system can be used as a real-time pre-alert system for nuclear power plant regarding tritium-in water environmental surveillance. The first detector module was characterized, commissioned and installed immediately after the discharge channel of the Arrocampo dam (Almaraz nuclear power plant, Spain) on the Tagus river. Due to the high sensitivity of the single detection modules, the system requires radioactive background mitigation techniques through the use of active and passive shielding. We have extrapolated a \emph{MDA} of 3.6\,kBq/L for a single module being this value limited by the cosmic background. The obtained value for a single module is already compatible with a real-time environmental surveillance and pre-alert system. Further optimization of the single-module sensitivity will imply the reduction of the number of modules and the cost of the detector system.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑