基于实时深度卷积神经网络的植入式心脏泵无传感器控制系统

A Sensorless Control System for an Implantable Heart Pump using a Real-time Deep Convolutional Neural Network

UNSW(新南威尔士大学);St Vincent’s Hospital(圣文森特医院);Victor Chang Cardiac Research Institute(维克多张心脏研究所)

Masoud Fetanat, Michael Stevens, Christopher Hayward, Nigel H. Lovell

摘要:

左心室辅助装置(LVAD)是一种机械泵,可用于支持心力衰竭(HF)患者,作为移植和目标治疗的桥梁。为了自动调节LVAD速度,需要设计生理控制系统,以响应各种临床情况下患者血液动力学的变化。这些控制系统需要来自心血管系统的压力反馈信号。然而,没有合适的长期植入式传感器可用。在这项研究中,提出了一种新的实时深度卷积神经网络(CNN),用于基于LVAD流量估计前负荷。采用无模型自适应控制(FFDL-MFAC)的全动态形式和所提出的前负荷估计器,开发了一种新的用于LVAD泵的无传感器自适应生理控制系统,以将患者的条件维持在安全的生理范围内。通过对100种不同患者状况的10倍交叉验证,对用于前负荷估计的CNN模型进行了训练和评价,并在6种不同患者场景的30种不同患者状况的新测试集上对所提出的无传感器控制系统进行了评估。所提出的前负荷估计器非常准确,相关系数为0.97,根均方误差为0.84 mmHg,再现性系数为1.56 mmHg,变异系数为14.44%,测试数据集的偏倚为0.29 mmHg。结果还表明,所提出的无传感器生理控制器的工作原理类似于基于预负荷的生理控制系统,用于LVAD,使用测量的预负荷,以防止心室抽吸和肺充血。该研究表明,LVAD可以适当地响应不断变化的患者状态和生理需求,而无需额外的压力或流量测量。

英文摘要:

Left ventricular assist devices (LVADs) are mechanical pumps, which can be used to support heart failure (HF) patients as bridge to transplant and destination therapy. To automatically adjust the LVAD speed, a physiological control system needs to be designed to respond to variations of patient hemodynamics across a variety of clinical scenarios. These control systems require pressure feedback signals from the cardiovascular system. However, there are no suitable long-term implantable sensors available. In this study, a novel real-time deep convolutional neural network (CNN) for estimation of preload based on the LVAD flow was proposed. A new sensorless adaptive physiological control system for an LVAD pump was developed using the full dynamic form of model free adaptive control (FFDL-MFAC) and the proposed preload estimator to maintain the patient conditions in safe physiological ranges. The CNN model for preload estimation was trained and evaluated through 10-fold cross validation on 100 different patient conditions and the proposed sensorless control system was assessed on a new testing set of 30 different patient conditions across six different patient scenarios. The proposed preload estimator was extremely accurate with a correlation coefficient of 0.97, root mean squared error of 0.84 mmHg, reproducibility coefficient of 1.56 mmHg, coefficient of variation of 14.44 %, and bias of 0.29 mmHg for the testing dataset. The results also indicate that the proposed sensorless physiological controller works similarly to the preload-based physiological control system for LVAD using measured preload to prevent ventricular suction and pulmonary congestion. This study shows that the LVADs can respond appropriately to changing patient states and physiological demands without the need for additional pressure or flow measurements.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑