对可植入型心律转换除颤器的模型检查
Model Checking Implantable Cardioverter Defibrillators
AI总结:
本文构建了心+ICD闭环的混合系统模型,证明其为STORMED系统,提出近似可达性可实现有限模拟,改进了验证流程,用于检查ICD闭环中的关键形式属性。
摘要:
心室颤动是心脏无序电兴奋导致血流不足的病理现象,通常数秒内致命。最常见的治疗方式是植入可植入型心律转换除颤器(ICD)。基于模型的验证可提供严格的安全性和有效性证明。本文构建了人体心脏+ICD闭环的混合系统模型,并证明其为STORMED系统,即一类允许有限仿真的最小化混合系统。通常无法计算仿真,但本文证明近似可达性可为STORMED系统提供有限仿真,从而改进现有验证流程。在过程中,证明了某些组合保持STORMED属性。因此,可以对ICD闭环中的关键形式属性进行模型检查,如延迟治疗、遗漏治疗或不恰当的治疗。本文结果具有理论意义,激励了STORMED系统的具体模型检查流程的创建。
英文摘要:
Ventricular Fibrillation is a disorganized electrical excitation of the heart that results in inadequate blood flow to the body. It usually ends in death within seconds. The most common way to treat the symptoms of fibrillation is to implant a medical device, known as an Implantable Cardioverter Defibrillator (ICD), in the patient's body. Model-based verification can supply rigorous proofs of safety and efficacy. In this paper, we build a hybrid system model of the human heart+ICD closed loop, and show it to be a STORMED system, a class of o-minimal hybrid systems that admit finite bisimulations. In general, it may not be possible to compute the bisimulation. We show that approximate reachability can yield a finite simulation for STORMED systems, which improves on the existing verification procedure. In the process, we show that certain compositions respect the STORMED property. Thus it is possible to model check important formal properties of ICDs in a closed loop with the heart, such as delayed therapy, missed therapy, or inappropriately administered therapy. The results of this paper are theoretical and motivate the creation of concrete model checking procedures for STORMED systems.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。