arXiv 2103.10193physics.ins-dethep-ex
收藏

用于测试JUNO的20'000个20英寸PMT的集装箱式设施

A container-based facility for testing 20'000 20-inch PMTs for JUNO

Björn Wonsak, Alexander Tietzsch, Tobias Sterr, Tobias Lachenmaier, Nikolay Anfimov, David Blum, Caren Hagner, Benedict Kaiser, David Meyhöfer, Alexander Olshev… 展开作者

Björn Wonsak, Alexander Tietzsch, Tobias Sterr, Tobias Lachenmaier, Nikolay Anfimov, David Blum, Caren Hagner, Benedict Kaiser, David Meyhöfer, Alexander Olshevskiy, Zhonghua Qin, Henning Rebber, Simon Reichert, Malte Stender, Wei Wang, Zhimin Wang

摘要:

JUNO实验的主要目标是确定中微子的质量排序。为了实现这一点,需要在$1\,$ MeV下至少达到$3\,\%$的非凡能量分辨率,为此JUNO探测器的所有部件都需要满足一定的质量标准。这一点特别关系到那些与探测器的能量分辨率有关的探测器,例如将在JUNO部署的光电倍增管。本文介绍了一个专门的光电倍增管质量测试设备的设置和性能,以检查和表征20英寸JUNO光电倍增管的性能。它的准工业规模和操作水平允许测试所有20000 PMT打算在JUNO实验中使用。与此PMT质量测试系统,几个关键特性,如暗计数率,峰谷比,光子探测效率,和定时分辨率已被确定在$1\times10^7$的工作增益和评估相对于JUNO的要求。测量条件和模式的光电倍增管,以及估计精度的个人光电倍增管参数的确定与系统。

英文摘要:

The main goal of the JUNO experiment is the determination of the neutrino mass ordering. To achieve this, an extraordinary energy resolution of at least $3\,\%$ at $1\,$MeV is required for which all parts of the JUNO detector need to meet certain quality criteria. This is relevant in particular for those which are related to the energy resolution of the detector, such as the photomultiplier tubes (PMTs) to be deployed in JUNO. This paper presents the setup and performance of a dedicated PMT mass testing facility to examine and characterize the performance of the 20-inch JUNO PMTs. Its quasi-industrial size and operation level allows to test all 20000 PMTs intended to be used in the JUNO experiment. With this PMT mass testing system, several key characteristics like dark count rate, peak-to-valley ratio, photon detection efficiency, and timing resolution have been determined at an operating gain of $1\times10^7$ and assessed with respect to the requirements of JUNO. Measurement conditions and modes for the PMTs as well as estimated accuracies for the determination of the individual PMT parameters with the system are presented as well.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑