具有宽带X射线谱的32 $\times$ 32 Ti/Au TES阵列的单像素性能
Single pixel performance of a 32$\times$32 Ti/Au TES array with broadband X-ray spectra
摘要:
我们正在开发一种千像素Ti/Au TES阵列,作为Athena X-IFU的备用选项。在这里,我们报告的单像素性能的32 $\times$ 32阵列中操作的频分复用(FDM)读出系统,偏置频率在1-5 MHz的范围内。我们已经通过$^{55}$ Fe放射源(发射5.9 keV的Mn-K $α$)和调制X射线源(MXS,提供5.4 keV的Cr-K $α$和8.0 keV的Cu-K $α$)测试了几种光子能量下的像素响应。首先,我们报告的程序用于执行探测器的能量标度校准,通常实现的校准精度优于$\sim$ 0.5 eV的5.4 - 8.9 keV的能量范围。然后,我们给出了在不同能量下测量的能量分辨率(最佳单像素性能:$Δ$ E $_{FWHM}$ = 2.40 $\pm$ 0.09 eV@5.4 keV; 2.53 $\pm$ 0.10 eV@5.9 keV; 2.78 $\pm$ 0.16 eV@8.0 keV),还研究了来自像素偏置频率和计数率的性能依赖性。由于长时间的背景测量($\sim$ 1天),我们最终还检测到了1.5 keV的Al-K $α$线,该线由实验装置内部的荧光产生。我们分析了这条线,以获得也在低能量($Δ$ E $_{FWHM}$ = 1.91 eV $\pm$ 0.21 eV@1.5 keV)下的单像素性能的第一评估,并评估在大能带(1.5 - 8.9 keV)中的检测器响应的线性。
英文摘要:
We are developing a kilo-pixels Ti/Au TES array as a backup option for Athena X-IFU. Here we report on single-pixel performance of a 32$\times$32 array operated in a Frequency Division Multiplexing (FDM) readout system, with bias frequencies in the range 1-5 MHz. We have tested the pixels response at several photon energies, by means of a $^{55}$Fe radioactive source (emitting Mn-K$α$ at 5.9 keV) and a Modulated X-ray Source (MXS, providing Cr-K$α$ at 5.4 keV and Cu-K$α$ at 8.0 keV). First, we report the procedure used to perform the detector energy scale calibration, usually achieving a calibration accuracy better than $\sim$ 0.5 eV in the 5.4 - 8.9 keV energy range. Then, we present the measured energy resolution at the different energies (best single pixel performance: $Δ$E$_{FWHM}$ = 2.40 $\pm$ 0.09 eV @ 5.4 keV; 2.53 $\pm$ 0.10 eV @ 5.9 keV; 2.78 $\pm$ 0.16 eV @ 8.0 keV), investigating also the performance dependency from the pixel bias frequency and the count rate. Thanks to long background measurements ($\sim$ 1 day), we finally detected also the Al-K$α$ line at 1.5 keV, generated by fluorescence inside the experimental setup. We analyzed this line to obtain a first assessment of the single-pixel performance also at low energy ($Δ$E$_{FWHM}$ = 1.91 eV $\pm$ 0.21 eV @ 1.5 keV), and to evaluate the linearity of the detector response in a large energy band (1.5 - 8.9 keV).
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。