具有宽带X射线谱的32 $\times$ 32 Ti/Au TES阵列的单像素性能

Single pixel performance of a 32$\times$32 Ti/Au TES array with broadband X-ray spectra

Matteo D'Andrea, Emanuele Taralli, Hiroki Akamatsu, Luciano Gottardi, Kenichiro Nagayoshi, Kevin Ravensberg, Marcel L. Ridder, Davide Vaccaro, Cor P. de Vries, … 展开作者

Matteo D'Andrea, Emanuele Taralli, Hiroki Akamatsu, Luciano Gottardi, Kenichiro Nagayoshi, Kevin Ravensberg, Marcel L. Ridder, Davide Vaccaro, Cor P. de Vries, Martin de Wit, Marcel P. Bruijn, Ruud W. M. Hoogeveen, Jian-Rong Gao

摘要:

我们正在开发一种千像素Ti/Au TES阵列,作为Athena X-IFU的备用选项。在这里,我们报告的单像素性能的32 $\times$ 32阵列中操作的频分复用(FDM)读出系统,偏置频率在1-5 MHz的范围内。我们已经通过$^{55}$ Fe放射源(发射5.9 keV的Mn-K $α$)和调制X射线源(MXS,提供5.4 keV的Cr-K $α$和8.0 keV的Cu-K $α$)测试了几种光子能量下的像素响应。首先,我们报告的程序用于执行探测器的能量标度校准,通常实现的校准精度优于$\sim$ 0.5 eV的5.4 - 8.9 keV的能量范围。然后,我们给出了在不同能量下测量的能量分辨率(最佳单像素性能:$Δ$ E $_{FWHM}$ = 2.40 $\pm$ 0.09 eV@5.4 keV; 2.53 $\pm$ 0.10 eV@5.9 keV; 2.78 $\pm$ 0.16 eV@8.0 keV),还研究了来自像素偏置频率和计数率的性能依赖性。由于长时间的背景测量($\sim$ 1天),我们最终还检测到了1.5 keV的Al-K $α$线,该线由实验装置内部的荧光产生。我们分析了这条线,以获得也在低能量($Δ$ E $_{FWHM}$ = 1.91 eV $\pm$ 0.21 eV@1.5 keV)下的单像素性能的第一评估,并评估在大能带(1.5 - 8.9 keV)中的检测器响应的线性。

英文摘要:

We are developing a kilo-pixels Ti/Au TES array as a backup option for Athena X-IFU. Here we report on single-pixel performance of a 32$\times$32 array operated in a Frequency Division Multiplexing (FDM) readout system, with bias frequencies in the range 1-5 MHz. We have tested the pixels response at several photon energies, by means of a $^{55}$Fe radioactive source (emitting Mn-K$α$ at 5.9 keV) and a Modulated X-ray Source (MXS, providing Cr-K$α$ at 5.4 keV and Cu-K$α$ at 8.0 keV). First, we report the procedure used to perform the detector energy scale calibration, usually achieving a calibration accuracy better than $\sim$ 0.5 eV in the 5.4 - 8.9 keV energy range. Then, we present the measured energy resolution at the different energies (best single pixel performance: $Δ$E$_{FWHM}$ = 2.40 $\pm$ 0.09 eV @ 5.4 keV; 2.53 $\pm$ 0.10 eV @ 5.9 keV; 2.78 $\pm$ 0.16 eV @ 8.0 keV), investigating also the performance dependency from the pixel bias frequency and the count rate. Thanks to long background measurements ($\sim$ 1 day), we finally detected also the Al-K$α$ line at 1.5 keV, generated by fluorescence inside the experimental setup. We analyzed this line to obtain a first assessment of the single-pixel performance also at low energy ($Δ$E$_{FWHM}$ = 1.91 eV $\pm$ 0.21 eV @ 1.5 keV), and to evaluate the linearity of the detector response in a large energy band (1.5 - 8.9 keV).

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑