可变形闪烁剂量计I:挑战和使用计算机视觉技术的实现

Deformable Scintillation Dosimeter I: Challenges and Implementation using Computer Vision Techniques

Emily Cloutier, Louis Archambault, Luc Beaulieu

摘要:

塑料闪烁探测器越来越多地用于测量放射治疗背景下的剂量分布。它们的水等效性,实时响应和高空间分辨率使其区别于传统的探测器,特别是在复杂的辐射几何形状中。它们的应用范围可以通过在模仿解剖学变化的可变形基质中嵌入生物传感器来进一步扩展。在这项工作中,我们的特点所产生的信号变化的平移和旋转的反射纤维相对于相机。提出了使用立体视觉技术和两个sCMOS补充CCD相机的校正。该研究扩展到一个原型的实时变形剂量计的情况下,包括一个阵列的19个闪烁。信号与角度的关系遵循高斯分布(FWHM = 52°),而来自径向位移的强度变化遵循平方反比定律。跟踪纤维的位置和角度能够校正这些空间依赖性。该检测系统的位置检测精度为0.008cm,精度为0.03cm。这导致角度测量的不确定度为2°。将剂量计移位$\pm$ 3 cm深度,导致相对于参考位置的相对强度为100 $\pm$ 10%(平均$\pm$标准差)。应用校正减少了变化,从而导致相对强度为100 $\pm$ 1%。同样,对于$\pm$ 3 cm的横向位移,校正后强度从98 $\pm$ 3%变为100 $\pm$ 1%。因此,对由对3D体积中的成像元件的输出进行成像的相机收集的信号的精确校正是可能的。这项工作铺平了道路,以实时闪烁器为基础的变形剂量计的发展。

英文摘要:

Plastic scintillation detectors are increasingly used to measure dose distributions in the context of radiotherapy treatments. Their water-equivalence, real-time response and high spatial resolution distinguish them from traditional detectors, especially in complex irradiation geometries. Their range of applications could be further extended by embedding scintillators in a deformable matrix mimicking anatomical changes. In this work, we characterized signal variations arising from the translation and rotation of scintillating fibers with respect to a camera. Corrections are proposed using stereo vision techniques and two sCMOS complementing a CCD camera. The study was extended to the case of a prototype real-time deformable dosimeter comprising an array of 19 scintillating. The signal to angle relationship follows a gaussian distribution (FWHM = 52°) whereas the intensity variation from radial displacement follows the inverse square law. Tracking the position and angle of the fibers enabled the correction of these spatial dependencies. The detecting system provides an accuracy and precision of respectively 0.008 cm and 0.03 cm on the position detection. This resulted in an uncertainty of 2° on the angle measurement. Displacing the dosimeter by $\pm$3 cm in depth resulted in relative intensities of 100$\pm$10% (mean $\pm$ standard deviation) to the reference position. Applying corrections reduced the variations thus resulting in relative intensities of 100$\pm$1%. Similarly, for lateral displacements of $\pm$3 cm, intensities went from 98$\pm$3% to 100$\pm$1% after the correction. Therefore, accurate correction of the signal collected by a camera imaging the output of scintillating elements in a 3D volume is possible. This work paves the way to the development of real-time scintillator-based deformable dosimeters.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑