MPA-MURA编码孔径准直器肺内污染成像系统的模拟
Simulation of an imaging system for internal contamination of lungs using MPA-MURA coded aperture collimator
摘要:
核事故中吸入的放射性核素常引起低活度的肺部内污染。平行孔成像系统广泛应用于医用γ相机中,虽然具有高分辨率和良好的图像质量,但由于其极低的检测效率,仍然难以获得吸入性肺部污染的图像。本文采用蒙特卡罗方法研究了MPA-MURA编码孔径准直器对肺内污染成像的影响。成像系统由一个成年男性肺模型,一个镶嵌,图案为中心,反对称MURA编码孔径准直器模型和CsI(Tl)探测器模型。采用MLEM解码算法重建内部污染图像,采用互补成像方法减少伪影。MPA-MURA编码孔径成像重建的I-131点源图像的半高宽为2.51 mm,简化呼吸道源图像的信噪比为3.98 dB。虽然MPA-MURA编码孔径成像的空间分辨率不如平行孔成像,但其探测效率比平行孔准直器成像高两个数量级。考虑到核事故引起的肺内污染的低放射性水平,PMA-MURA编码孔径成像在肺污染成像方面具有重要的发展潜力。
英文摘要:
The nuclides inhaled during nuclear accidents usually cause internal contamination of the lungs with low activity. Although a parallel-hole imaging system, which is widely used in medical gamma cameras, has a high resolution and good image quality, owing to its extremely low detection efficiency, it remains difficult to obtain images of inhaled lung contamination. In this study, the Monte Carlo method was used to study the internal lung contamination imaging using the MPA-MURA coded-aperture collimator. The imaging system consisted of an adult male lung model, with a mosaicked, pattern-centered, and anti-symmetric MURA coded-aperture collimator model and a CsI(Tl) detector model. The MLEM decoding algorithm was used to reconstruct the internal contamination image, and the complementary imaging method was used to reduce the number of artifacts. The full width at half maximum of the I-131 point source image reconstructed by the mosaicked, pattern-centered, and anti-symmetric Modified uniformly redundant array (MPA-MURA) coded-aperture imaging reached 2.51 mm, and the signal-to-noise ratio of the simplified respiratory tract source (I-131) image reconstructed through MPA-MURA coded-aperture imaging was 3.98 dB. Although the spatial resolution of MPA-MURA coded aperture imaging is not as good as that of parallel-hole imaging, the detection efficiency of PMA-MURA coded-aperture imaging is two orders of magnitude higher than that of parallel hole collimator imaging. Considering the low activity level of internal lung contamination caused by nuclear accidents, PMA-MURA coded-aperture imaging has significant potential for the development of lung contamination imaging.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。