Imaging of Spatially Extended Hot Spots with Coded Apertures for Intra-operative Nuclear Medicine Applications
摘要:
编码孔径成像在效率和视场(FoV)方面超越了传统准直器的平面成像。我们给出了探测均匀扩展的99 mTc和57 Co热点发射的141 keV和122 keV γ光子的实验结果以及对均匀和正常扩展的99 mTc热点的模拟。这些结果证明,该方法可用于术中成像的放射性示踪前哨淋巴结和甲状腺残留物。使用两个伽马相机的设置进行研究,每个相机由位于CdTe探测器顶部的等级19的修改的均匀冗余阵列(MURA)的编码孔径(或掩模)组成。探测器像素间距为350 μm,其有效面积为4.4 × 4.4 cm 2,而掩模元件尺寸为1.7 mm。可探测的光子能量范围从15 keV到200 keV,能量分辨率为3-4 keV FWHM。利用三角测量来估计系统FoV内的放射性斑点的3D空间坐标。两个扩展源,具有均匀分布的活动(直径分别为11和24毫米),位于16厘米的系统和3cm的中心之间的距离,可以解决和定位精度优于5%。结果表明,空间扩展源的估算位置在其体积大小范围内,即使是低放射性水平(如30 MBq)的邻近源,也可以在约3秒的计数时间内清楚地区分
英文摘要:
Coded aperture imaging transcends planar imaging with conventional collimators in efficiency and Field of View (FoV). We present experimental results for the detection of 141keV and 122keV γ-photons emitted by uniformly extended 99mTc and 57Co hot-spots along with simulations of uniformly and normally extended 99mTc hot-spots. These results prove that the method can be used for intra-operative imaging of radio-traced sentinel nodes and thyroid remnants. The study is performed using a setup of two gamma cameras, each consisting of a coded-aperture (or mask) of Modified Uniformly Redundant Array (MURA) of rank 19 positioned on top of a CdTe detector. The detector pixel pitch is 350 μm and its active area is 4.4x4.4 cm2, while the mask element size is 1.7mm. The detectable photon energy ranges from 15 keV up to 200 keV with an energy resolution of 3-4 keV FWHM. Triangulation is exploited to estimate the 3D spatial coordinates of the radioactive spots within the system FoV. Two extended sources, with uniform distributed activity (11 and 24 mm in diameter, respectively), positioned at 16cm from the system and with 3cm distance between their centers, can be resolved and localized with accuracy better than 5%. The results indicate that the estimated positions of spatially extended sources lay within their volume size and that neighboring sources, even with a low level of radioactivity, such as 30 MBq, can be clearly distinguished with counting time about 3 seconds
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。