A plan quality control method of treatment planning for Gamma Knife radiosurgery

Tonghe Wang, Matt D. Giles, Elizabeth Butker, Matthew C. Walb, Xiaofeng Yang, Tian Liu, Shannon Kahn, Zhen Tian

摘要:

由于需要调整的变量很多,传统的手动伽玛刀(GK)放射外科正向计划非常复杂和繁琐。由此产生的计划质量在很大程度上取决于规划者的技能,经验和投入的努力,并在案件,规划者和机构之间有很大的差异。GK计划的质量控制需要始终如一地为每位患者提供高质量的计划。在本研究中,我们提出了一个质量控制的方法,通过建立一个高质量的GK计划数据库的GK计划。通过目标体积、目标形状复杂度以及目标与邻近器官的空间关系来描述患者解剖结构,这决定了GK计划的难度。计划的质量进行了评估,使用目标覆盖率,选择性,中等剂量溢出,最大剂量0.1毫升的脑干,同侧耳蜗的平均剂量,和光束的时间。当创建新计划时,将从数据库中识别出具有最相似靶体积大小和形状复杂性的高质量计划。根据脑干和耳蜗的重叠体积直方图,建立了脑干和耳蜗剂量预测模型。识别的参考计划和预测的器官剂量将帮助计划者相应地做出质量控制决策。为了验证这种方法,我们已经建立了一个数据库前庭神经鞘瘤,这被认为是具有挑战性的GK规划,由于不规则形状的目标和它的接近脑干和耳蜗。测试了5个案例,其中1个案例被认为是高质量的,4个案例的计划质量低于预测。这四个病例经过重新规划,得到了实质性的改善。我们的研究结果证明了我们提出的质量控制方法的有效性。该方法也可作为一种计划质量预测方法,以促进GK放射外科自动治疗计划的发展。

英文摘要:

With many variables to adjust, conventional manual forward planning for Gamma Knife (GK) radiosurgery is very complicated and cumbersome. The resulting plan quality heavily depends on planners skills, experiences and devoted efforts, and varies significantly among cases, planners, and institutions. Quality control for GK planning is desired to consistently provide high-quality plan to each patient. In this study, we proposed a quality control method for GK planning by building a database of high-quality GK plans. Patient anatomy was described by target volume, target shape complexity, and spatial relationship between target and nearby organs, which determine GK planning difficulty level. Plan quality was evaluated using target coverage, selectivity, intermediate dose spillage, maximum dose to 0.1 cc of brainstem, mean dose of ipsilateral cochlea, and beam-on time. When a new plan is created, a high-quality plan that has the most similar target volume size and shape complexity will be identified from the database. A model has also been built to predict the dose to brainstem and cochlea based on their overlap volume histograms. The identified reference plan and the predicted organ dose will help planners to make quality control decisions accordingly. To validate this method, we have built a database for vestibular schwannoma, which are considered to be challenging for GK planning due to the irregularly-shaped target and its proximity to brainstem and cochlea. Five cases were tested, among which one case was considered to be of high quality and four cases had a lower plan quality than prediction. These four cases were replanned and got substantially improved. Our results have demonstrated the efficacy of our proposed quality control method. This method may also be used as a plan quality prediction method to facilitate the development of automatic treatment planning for GK radiosurgery.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑