arXivDaily arXiv每日学术速递 周一至周五更新
arXiv周末暂无论文更新,休息一下吧,周末愉快~~
arXiv 2008.06029eess.IVcs.CVcs.LGeess.SPphysics.med-ph
收藏

高加速MRI中物理引导神经网络的多模板自监督学习

Multi-Mask Self-Supervised Learning for Physics-Guided Neural Networks in Highly Accelerated MRI

University of Minnesota(明尼苏达大学)

Burhaneddin Yaman, Hongyi Gu, Seyed Amir Hossein Hosseini, Omer Burak Demirel, Steen Moeller, Jutta Ellermann, Kâmil Uğurbil, Mehmet Akçakaya

摘要:

自监督学习由于其在没有完全采样数据的情况下训练深度学习MRI重建方法的能力而显示出巨大的前景。当前用于物理引导重建网络的自监督学习方法将采集的欠采样数据分成两个不相交的集合,其中一个用于展开网络中的数据一致性(DC),另一个用于定义训练损失。在这项研究中,我们提出了一种改进的自监督学习策略,更有效地使用所获得的数据训练物理指导的重建网络,而无需完全采样数据的数据库。所提出的通过数据欠采样的多掩码自监督学习(SSDU)对所获取的测量值应用保持掩码操作,以将其分成用于每个训练样本的多对不相交的集合,同时使用这些对中的一个用于DC单元,另一个用于定义损失,从而更有效地使用欠采样数据。多掩模SSDU应用于全采样3D膝关节和前瞻性欠采样3D脑部MRI数据集,用于各种加速度和模式,并与CG-SENSE和单掩模SSDU DL-MRI以及监督DL-MRI(当全采样数据可用时)进行比较。膝关节MRI的结果表明,建议的多掩模SSDU优于SSDU,并与监督DL-MRI的性能接近。一项临床阅片师研究进一步将多掩模SSDU在SNR和混叠伪影方面的排名高于监督DL-MRI。脑MRI结果表明,多掩模SSDU实现更好的重建质量相比,SSDU。读者研究表明,与R =8时的单掩模SSDU以及R=2时的CG-SENSE相比,R = 8时的多掩模SSDU显著改善了重建。

英文摘要:

Self-supervised learning has shown great promise due to its capability to train deep learning MRI reconstruction methods without fully-sampled data. Current self-supervised learning methods for physics-guided reconstruction networks split acquired undersampled data into two disjoint sets, where one is used for data consistency (DC) in the unrolled network and the other to define the training loss. In this study, we propose an improved self-supervised learning strategy that more efficiently uses the acquired data to train a physics-guided reconstruction network without a database of fully-sampled data. The proposed multi-mask self-supervised learning via data undersampling (SSDU) applies a hold-out masking operation on acquired measurements to split it into multiple pairs of disjoint sets for each training sample, while using one of these pairs for DC units and the other for defining loss, thereby more efficiently using the undersampled data. Multi-mask SSDU is applied on fully-sampled 3D knee and prospectively undersampled 3D brain MRI datasets, for various acceleration rates and patterns, and compared to CG-SENSE and single-mask SSDU DL-MRI, as well as supervised DL-MRI when fully-sampled data is available. Results on knee MRI show that the proposed multi-mask SSDU outperforms SSDU and performs closely with supervised DL-MRI. A clinical reader study further ranks the multi-mask SSDU higher than supervised DL-MRI in terms of SNR and aliasing artifacts. Results on brain MRI show that multi-mask SSDU achieves better reconstruction quality compared to SSDU. Reader study demonstrates that multi-mask SSDU at R=8 significantly improves reconstruction compared to single-mask SSDU at R=8, as well as CG-SENSE at R=2.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑