连续工作的干涉仪由环境湿度、温度和压力变化引起的相位漂移的补偿
Compensation of phase drifts caused by ambient humidity, temperature and pressure changes for continuously operating interferometers
摘要:
聚变实验在很大程度上依赖于通过干涉测量法测量线积分电子密度,以进行密度反馈控制。近年来,放电长度急剧增加,并继续上升,导致环境引起的相位漂移成为一个越来越令人担忧的问题,因为它们伪造干涉仪的密度测量。特别是在较大的托卡马克中,由于光路长度的不受控制的变化而导致的密度控制的损失可能具有灾难性的结果。在大型诊断/实验大厅中控制环境参数是昂贵的,并且有时是不可行的,并且在某些情况下不能对现有机器进行改造。在本报告中,我们提出了一个非常便宜的(约。100欧元),易于改装,实时相位补偿方案的干涉仪测量色散介质在长时间尺度。该方法不限于融合,但可以应用于任何连续测量干涉仪测量色散介质。它已成功地应用于Wendelstein 7-X密度反馈干涉仪。
英文摘要:
Fusion experiments rely heavily on the measurement of the line-integrated electron density by interferometry for density feed-back control. In recent years the discharge length has increased dramatically and is continuing to rise, resulting in environmentally induced phase drifts to become an increasingly worrisome subject, since they falsify the interferometer's measurement of the density. Especially in larger Tokamaks the loss of density control due to uncontrolled changes in the optical path length can have a disastrous outcome. The control of environmental parameters in large diagnostic/experimental halls is costly and sometimes infeasible and in some cases cannot be retro-fitted to an existing machine. In this report we present a very cheap (ca. 100 EUR), easily retro-fitted, real-time capable phase compensation scheme for interferometers measuring dispersive media over long time scales. The method is not limited to fusion, but can be applied to any continuously measuring interferometer measuring a dispersive medium. It has been successfully applied to the Wendelstein 7-X density feed-back interferometer.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。