Herschel-SPIRE傅里叶变换光谱仪扩展源定标的修正

Correcting the extended-source calibration for the Herschel-SPIRE Fourier-Transform Spectrometer

Ivan Valtchanov, R. Hopwood, G. Bendo, C. Benson, L. Conversi, T. Fulton, M. J. Griffin, T. Joubaud, T. Lim, N. Lu, N. Marchili, G. Makiwa, R. A. Meyer, D. A. N… 展开作者

Ivan Valtchanov, R. Hopwood, G. Bendo, C. Benson, L. Conversi, T. Fulton, M. J. Griffin, T. Joubaud, T. Lim, N. Lu, N. Marchili, G. Makiwa, R. A. Meyer, D. A. Naylor, C. North, A. Papageorgiou, C. Pearson, E. T. Polehampton, J. Scott, B. Schulz, L. D. Spencer, M. H. D. van der Wiel, R. Wu

摘要:

我们描述了一个更新的Herschel-SPIRE傅里叶变换光谱仪(FTS)校准扩展源,其中包括一个校正的频率相关的远场喇叭的效率,$η_\mathrm{FF}$。无论光谱分辨率如何,这种显著的校正都会影响稀疏或映射模式下的所有FTS扩展源校准光谱。谱线通量和连续能级在光谱仪长波段(SLW,447-1018 GHz; 671-294 $μ$ m)被低估了1.3-2倍,在光谱仪短波段(SSW,944-1568 GHz; 318-191 $μ$ m)被低估了1.4-1.5倍。该纠正在FTS管道版本14.1中实施,自2017年2月以来,在SPIRE手册中也进行了描述。在此管道版本之前,基于扩展源校准光谱的研究应使用赫歇尔科学档案中现有的产品进行批判性的重新考虑。一旦扩展源校准光谱被校正为$η_\mathrm{FF}$,合成光度学和来自SPIRE光度计图的宽带强度在2-4%内一致-与点源校准光谱和来自点源校准图的光度学的比较相似的水平。FTS的两种校准方案现在是自洽的:校正的扩展源和点源校准光谱之间的转换可以通过光束立体角和考虑衍射损失的增益校正来实现。

英文摘要:

We describe an update to the Herschel-SPIRE Fourier-Transform Spectrometer (FTS) calibration for extended sources, which incorporates a correction for the frequency-dependent far-field feedhorn efficiency, $η_\mathrm{FF}$. This significant correction affects all FTS extended-source calibrated spectra in sparse or mapping mode, regardless of the spectral resolution. Line fluxes and continuum levels are underestimated by factors of 1.3-2 in the Spectrometer Long-Wavelength band (SLW, 447-1018 GHz; 671-294 $μ$m) and 1.4-1.5 in the Spectrometer Short-Wavelength band (SSW, 944-1568 GHz; 318-191 $μ$m). The correction was implemented in the FTS pipeline version 14.1 and has also been described in the SPIRE Handbook since Feb 2017. Studies based on extended-source calibrated spectra produced prior to this pipeline version should be critically reconsidered using the current products available in the Herschel Science Archive. Once the extended-source calibrated spectra are corrected for $η_\mathrm{FF}$, the synthetic photometry and the broadband intensities from SPIRE photometer maps agree within 2-4% -- similar levels to the comparison of point-source calibrated spectra and photometry from point-source calibrated maps. The two calibration schemes for the FTS are now self-consistent: the conversion between the corrected extended-source and point-source calibrated spectra can be achieved with the beam solid angle and a gain correction that accounts for the diffraction loss.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑