在LHC运行1期间Atlas Tile量热计的Laser校准

The Laser calibration of the Atlas Tile Calorimeter during the LHC run 1

ATLAS Tile Calorimeter system

摘要:

本文介绍了在LHC第一次运行中使用的阿特拉斯强子瓦量热计的激光校准系统。首先,系统相关的读出电子学的稳定性进行了研究。该方法稳定,变异小于0.6%。然后,发展的方法来计算校准常数,以校正的增益的量热计光电倍增管的变化,进行了说明。这些常数的测定具有0.3%的统计不确定度和0.2%的热量计的中心部分和0.5%的端盖的系统不确定度。最后介绍了利用激光系统对瓦式量热仪定时错位的检测和校正。

英文摘要:

This article describes the Laser calibration system of the Atlas hadronic Tile Calorimeter that has been used during the run 1 of the LHC. First, the stability of the system associated readout electronics is studied. It is found to be stable with variations smaller than 0.6 %. Then, the method developed to compute the calibration constants, to correct for the variations of the gain of the calorimeter photomultipliers, is described. These constants were determined with a statistical uncertainty of 0.3 % and a systematic uncertainty of 0.2 % for the central part of the calorimeter and 0.5 % for the end-caps. Finally, the detection and correction of timing mis-configuration of the Tile Calorimeter using the Laser system are also presented.