用于4.2 ~ 300 K磁阻尼测量的宽带铁磁共振勺形探头

A broadband Ferromagnetic Resonance dipper probe for magnetic damping measurements from 4.2 K to 300 K

Shikun He, Christos Panagopoulos

摘要:

描述了一种用于4.2K至室温宽频带铁磁共振(FMR)的勺形探头。该装置基于接地共面波导的双端口传输微波信号测量。波导产生微波场并记录样品响应。采用三级勺式设计,控温快速稳定。在液氦温度下,FMR引起的温度变化在毫开尔文范围内。我们还设计了一种新型的FMR探头与弹簧加载的样品架。与普通FMR磁头相比,实现了改善的信噪比和稳定性。使用超导矢量磁体,我们证明吉尔伯特阻尼测量两个薄膜样品使用矢量网络分析仪与频率高达26 GHz:1)坡莫合金膜的5 nm厚和2)CoFeB膜的1.5 nm厚。在平行和垂直于薄膜平面的外加磁场下进行了实验。

英文摘要:

A dipper probe for broadband Ferromagnetic Resonance (FMR) operating from 4.2 K to room temperature is described. The apparatus is based on a 2-port transmitted microwave signal measurement with a grounded coplanar waveguide. The waveguide generates a microwave field and records the sample response. A 3-stage dipper design is adopted for fast and stable temperature control. The temperature variation due to FMR is in the milli-Kelvin range at liquid helium temperature. We also designed a novel FMR probe head with a spring-loaded sample holder. Improved signal-to-noise ratio and stability compared to a common FMR head are achieved. Using a superconducting vector magnet we demonstrate Gilbert damping measurements on two thin film samples using a vector network analyzer with frequency up to 26 GHz: 1) A Permalloy film of 5 nm thickness and 2) a CoFeB film of 1.5 nm thickness. Experiments were performed with the applied magnetic field parallel and perpendicular to the film plane.