arXivDaily arXiv每日学术速递 周一至周五更新
感恩节,周五至周末无更新,大家顺遂。周一再来吧
arXiv 1503.03932physics.med-ph
收藏

使用机器日志数据的体积调制弧治疗(VMAT)输送的统计质量控制

Statistical quality control for volumetric modulated arc therapy (VMAT) delivery using machine log data

Kwang-Ho Cheong, Me-Yeon Lee, Sei-Kwon Kang, Jai-Woong Yoon, Soah Park, Taejin Hwang, Haeyoung Kim, Kyoung Ju Kim, Tae Jin Han, Hoonsik Bae

摘要:

本研究的目的是使用机器日志数据建立用于监测体积调制弧治疗(VMAT)输送误差的统计质量控制。Eclipse和Clinac iX直线加速器与RapidArc系统(Varian Medical Systems,Palo Alto,USA)用于交付VMAT计划。在输送RapidArc射野期间,机器确定输送的电机单位(MU)和机架角度位置精度,并在完成RapidArc输送后在控制台监视器上显示MU(sigma_MU;剂量测定误差)和机架角度(sigma_GA;几何误差)的标准偏差。在本研究中,首先分析日志数据以确认其有效性和可用性;然后,应用统计过程控制(SPC)及时监测所有RapidArc射野的sigma_MU和sigma_GA:99名患者共195个弧射野。所有射野的sigma_MU和sigma_GA测定两次,即首先在患者特定计划QA期间测定,然后在首次治疗期间再次测定。无论治疗部位如何,sigma_MU和sigma_GA时间序列都非常稳定;然而,sigma_GA强烈依赖于机架旋转速度。立体定向体部放射治疗(SBRT)的RapidArc输送的sigma_GA小于典型VMAT。因此,SPC分别适用于SBRT病例和一般病例。此外,机架旋转的电位计的精度是重要的,因为sigma_GA可以由于其条件而显著改变。通过将SPC应用于sigma_MU和sigma_GA,可以有效地监控交货错误。然而,SPC的上限和下限需要在充分了解机器和日志数据的情况下仔细确定。

英文摘要:

The aim of this study is to set up statistical quality control for monitoring of volumetric modulated arc therapy (VMAT) delivery error using machine log data. Eclipse and Clinac iX linac with the RapidArc system (Varian Medical Systems, Palo Alto, USA) is used for delivery of the VMAT plan. During the delivery of the RapidArc fields, the machine determines the delivered motor units (MUs) and gantry angle position accuracy and the standard deviations of MU (sigma_MU; dosimetric error) and gantry angle (sigma_GA; geometric error) are displayed on the console monitor after completion of the RapidArc delivery. In the present study, first, the log data was analyzed to confirm its validity and usability; then, statistical process control (SPC) was applied to monitor the sigma_MU and sigma_GA in a timely manner for all RapidArc fields: a total of 195 arc fields for 99 patients. The sigma_MU and sigma_GA were determined twice for all fields, that is, first during the patient-specific plan QA and then again during the first treatment. The sigma_MU and sigma_GA time series were quite stable irrespective of the treatment site; however, the sigma_GA was strongly dependent on the gantry rotation speed. The sigma_GA of the RapidArc delivery for stereotactic body radiation therapy (SBRT) was smaller than that for the typical VMAT. Therefore, SPC was applied for SBRT cases and general cases respectively. Moreover, the accuracy of the potential meter of gantry rotation is important, since the sigma_GA can be dramatically changed due to its condition. By applying SPC to the sigma_MU and sigma_GA, we could monitor the delivery error efficiently. However, the upper and lower limits of SPC need to be determined carefully with full knowledge of the machine and log data.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑