arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 1405.3125physics.flu-dyn
收藏

FDA基准喷管模型壁面剪切应力和不确定性量化的时间分辨粒子图像速度测量

Time-resolved particle image velocimetry measurements with wall shear stress and uncertainty quantification for the FDA benchmark nozzle model

Jaime S. Raben, Prasanna Hariharan, Ronald Robinson, Richard Malinauskas, Pavlos P. Vlachos

摘要:

我们提出了使用先进的粒子图像测速(PIV)处理和后处理技术对医疗器械的计算流体动力学(CFD)分析的基准实验数据进行验证。这项工作是之前FDA赞助的多实验室研究的延伸,该研究使用了一种模仿几何形状的医疗器械,称为FDA基准喷嘴模型。在喷管喉部雷诺数为500、2,000、5,000和8,000时,在模型的五个重叠区域进行了时间分辨PIV分析。与先前的研究相比,图像的空间分辨率增加了两倍。使用总体相关、动态范围增强和相位相关处理数据,以提高信噪比和测量精度,并解析具有大速度范围和梯度的流动区域,这是许多血液接触医疗器械的典型特征。使用径向基函数(RBF)计算与器械安全性相关的参数,包括壁面和整体流中的剪切应力,以提高准确度。根据PIV测量结果计算了现场空间分辨压力分布、雷诺应力和能量耗散率。速度测量的不确定性估计直接从PIV相关平面,并在每个测量位置的壁面剪应力的不确定性分析,使用蒙特卡罗模型。(摘要截短)

英文摘要:

We present validation of benchmark experimental data for computational fluid dynamics (CFD) analyses of medical devices using advanced Particle Image Velocimetry (PIV) processing and post-processing techniques. This work is an extension of a previous FDA-sponsored multi-laboratory study, which used a medical device mimicking geometry referred to as the FDA benchmark nozzle model. Time-resolved PIV analysis was performed in five overlapping regions of the model for Reynolds numbers in the nozzle throat of 500, 2,000, 5,000, and 8,000. Images included a two-fold increase in spatial resolution in comparison to the previous study. Data was processed using ensemble correlation, dynamic range enhancement, and phase correlations to increase signal-to-noise ratios and measurement accuracy, and to resolve flow regions with large velocity ranges and gradients, which is typical of many blood-contacting medical devices. Parameters relevant to device safety, including shear stress at the wall and in bulk flow, were computed using radial basis functions (RBF) to improve accuracy. In-field spatially resolved pressure distributions, Reynolds stresses and energy dissipation rates were computed from PIV measurements. Velocity measurement uncertainty was estimated directly from the PIV correlation plane, and uncertainty analysis for wall shear stress at each measurement location was performed using a Monte Carlo model. (truncated abstract)

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑