arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 1106.0568nucl-ex
收藏

寻找无中微子双β衰变中大TeO2晶体作为低温测辐射热计的性能

Performance of a large TeO2 crystal as a cryogenic bolometer in searching for neutrinoless double beta decay

L. Cardani, L. Gironi, J. W. Beeman, I. Dafinei, Z. Ge, G. Pessina, S. Pirro, Y. Zhu

摘要:

测辐射热计是研究无中微子双β衰变的理想装置。增大单个探测器的质量将简化大型实验的构造,但也将减少由位于表面附近的α发射体引起的每单位质量的背景以及由外部和内部伽马引起的背景。我们提出了非常有前途的结果,获得了2.13公斤TeO2晶体。这个测辐射热计在位于格兰萨索国家实验室地下深处的稀释冰箱中冷却到10.5 mK的温度,是有史以来最大的热探测器。探测器的能量分辨率范围从3.9 keV(在145 keV)到7.8 keV(在2615伽马线的208 Tl)FWHM。我们讨论了每单位质量的背景,可以实现增加测辐射热计的质量减少。

英文摘要:

Bolometers are ideal devices in the search for neutrinoless Double Beta Decay. Enlarging the mass of individual detectors would simplify the construction of a large experiment, but would also decrease the background per unit mass induced by alpha-emitters located close to the surfaces and background arising from external and internal gamma's. We present the very promising results obtained with a 2.13 kg TeO2 crystal. This bolometer, cooled down to a temperature of 10.5 mK in a dilution refrigerator located deep underground in the Gran Sasso National Laboratories, represents the largest thermal detector ever operated. The detector exhibited an energy resolution spanning a range from 3.9 keV (at 145 keV) to 7.8 keV (at the 2615 gamma-line of 208Tl) FWHM. We discuss the decrease in the background per unit mass that can be achieved increasing the mass of a bolometer.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑