The CUORE experiment will search for neutrinoless double beta decay (0nDBD) of 130Te using an array of 988 TeO_2 bolometers operated at 10 mK in the Laboratori Nazionali del Gran Sasso (Italy). The detector is housed in a large cryogen-free cryostat cooled by pulse tubes and a high-power dilution refrigerator. The TeO_2 bolometers measure the event energies, and a precise and reliable energy calibration is critical for the successful identification of candidate 0nDBD and background events. The detector calibration system under development is based on the insertion of 12 gamma-sources that are able to move under their own weight through a set of guide tubes that route them from deployment boxes on the 300K flange down into position in the detector region inside the cryostat. The CUORE experiment poses stringent requirements on the maximum heat load on the cryostat, material radiopurity, contamination risk and the ability to fully retract the sources during normal data taking. Together with the integration into a unique cryostat, this requires careful design and unconventional solutions. We present the design, challenges, and expected performance of this low-temperature energy calibration system.
The Cryogenic Underground Observatory for Rare Events (CUORE) is a ton-scale cryogenic experiment designed to search for neutrinoless double-beta decay of $^{130}$Te and other rare events. The CUORE detector consists of 988 TeO$_2$ bolometers operated underground at 10 mK in a dilution refrigerator at the Laboratori Nazionali del Gran Sasso. Candidate events are identified through a precise measurement of their energy. The absolute energy response of the detectors is established by the regular calibration of each individual bolometer using gamma sources. The close-packed configuration of the C…
We collected 19.4 days of data from four 750 g TeO2 bolometers, and in three of them we were able to set the energy threshold around 3 keV using a new analysis technique. We found a background rate ranging from 25 cpd/keV/kg at 3 keV to 2 cpd/keV/kg at 25 keV, and a peak at 4.7 keV. The origin of this peak is presently unknown, but its presence is confirmed by a reanalysis of 62.7 kg.days of data from the finished CUORICINO experiment. Finally, we report the expected sensitivities of the CUORE0 (52 bolometers) and CUORE (988 bolometers) experiments to a WIMP annual modulation signal.
Cryogenic bolometers, with their excellent energy resolution, flexibility in material, and availability in high purity, are excellent detectors for the search for neutrinoless double beta decay. Kilogram-size single crystals of TeO_2 are utilized in CUORICINO for an array with a total detector mass of 40.7 kg. CUORICINO currently sets the most stringent limit on the halflife of Te-130 of T > 2.4x10^{24} yr (90% C.L.), corresponding to a limit on the effective Majorana neutrino mass in the range of < 0.2-0.9 eV. Based on technology developed for CUORICINO and its predecessors, CUORE is a next-g…
CUORE - the Cryogenic Underground Observatory for Rare Events - is an experiment searching for the neutrinoless double-beta ($0νββ$) decay of $^{130}$Te with an array of 988 TeO$_2$ crystals operated as bolometers at $\sim$10 mK in a large dilution refrigerator. With this detector, we aim for a $^{130}$Te $0νββ$ decay half-life sensitivity of $9\times10^{25}$ y with 5 y of live time, and a background index of $\lesssim 10^{-2}$ counts/keV/kg/y. Making an effort to maintain radiopurity by minimizing the bolometers' exposure to radon gas during their installation in the cryostat, we perform all …
Understanding the nature of neutrino masses will require physics beyond the long-standing Standard Model of particle physics. Neutrinoless double beta decay experiments like the Cryogenic Underground Observatory for Rare Events (CUORE) are uniquely suited for probing the remaining mysteries of neutrino mass, particularly the question of the neutrino's Majorana nature. CUORE will be a next-generation experiment at the Laboratori Nazionali del Gran Sasso in Italy; it will consist of an array of 988 TeO2 detector crystals operated at ~10 mK, following the bolometric technique established by the C…
The external shell of the CUORE cryostat is a large cryogen-free system designed to host the dilution refrigerator and the bolometers of the CUORE experiment in a low radioactivity environment. The three vessels that form the outer shell were produced and delivered to the Gran Sasso underground Laboratories in July 2012. In this paper, we describe the production techniques and the validation tests done at the production site in 2012.
CUORE is a 741 kg array of TeO2 bolometers that will search for the neutrinoless double beta decay of 130Te. The detector is being constructed at the Laboratori Nazionali del Gran Sasso in Italy, where it will begin taking data in 2015. The CUORE cryostat will cool several metric tonnes of material to below 1 K and the CUORE detector itself will operate at a typical temperature of 10 mK. At this temperature, the CUORE detector will be the coldest contiguous cubic meter in the known Universe.
CUORE is a proposed tightly packed array of 1000 TeO_{2} bolometers, each being a cube 5 cm on a side with a mass of 750 gms. The array consists of 25 vertical towers, arranged in a square, of 5 towers by 5 towers, each containing 10 layers of 4 crystals. The design of the detector is optimized for ultralow- background searches for neutrinoless double beta decay of ^{130}Te (33.8% abundance), cold dark matter, solar axions, and rare nuclear decays. A preliminary experiment involving 20 crystals of various sizes (MIBETA) has been completed, and a single CUORE tower is being constructed as a sma…
CUORE, an array of 988 TeO$_2$ bolometers, is about to be one of the most sensitive experiments searching for neutrinoless double-beta decay. Its sensitivity could be further improved by removing the background from $α$ radioactivity. A few years ago it has been pointed out that the signal from $β$s can be tagged by detecting the emitted Cherenkov light, which is not produced by $α$s. In this paper we confirm this possibility. For the first time we measured the Cherenkov light emitted by a CUORE crystal, and found it to be 100 eV at the $Q$-value of the decay. To completely reject the $α$ back…
Among the detectors used for rare event searches, such as neutrinoless Double Beta Decay (0$ν$DBD) and Dark Matter experiments, bolometers are very promising because of their favorable properties (excellent energy resolution, high detector efficiency, a wide choice of different materials used as absorber, ...). However, up to now, the actual interesting possibility to identify the interacting particle, and thus to greatly reduce the background, can be fulfilled only with a double read-out (i.e. the simultaneous and independent read out of heat and scintillation light or heat and ionization). T…
TeO2 bolometers have been used for many years to search for neutrinoless double beta decay in 130-Te. CUORE, a tonne-scale TeO2 detector array, recently published the most sensitive limit on the half-life, $T_{1/2}^{0ν} > 1.5 \times 10^{25}\,$yr, which corresponds to an upper bound of $140-400$~meV on the effective Majorana mass of the neutrino. While it makes CUORE a world-leading experiment looking for neutrinoless double beta decay, it is not the only study that CUORE will contribute to in the field of nuclear and particle physics. As already done over the years with many small-scale experi…
For the first time, a cadmium tungstate crystal scintillator enriched in $^{116}$Cd has been succesfully tested as a scintillating bolometer. The measurement was performed above ground at a temperature of 18 mK. The crystal mass was 34.5 g and the enrichment level ~82 %. Despite a substantial pile-up effect due to above-ground operation, the detector demonstrated a high energy resolution (2-7 keV FWHM in 0.2-2.6 MeV $γ$ energy range), a powerful particle identification capability and a high level of internal radiopurity. These results prove that cadmium tungstate is an extremely promising dete…
We have developed and tested a new way of coupling bolometric light detectors to scintillating crystal bolometers based upon simply resting the light detector on the crystal surface, held in position only by gravity. This straightforward mounting results in three important improvements: (1) it decreases the amount of non-active materials needed to assemble the detector, (2) it substantially increases the light collection efficiency by minimizing the light losses induced by the mounting structure, and (3) it enhances the thermal signal induced in the light detector thanks to the extremely weak …
CUORE is a tonne-scale cryogenic detector operating at the Laboratori Nazionali del Gran Sasso (LNGS) that uses tellurium dioxide bolometers to search for neutrinoless double-beta decay of $^{130}$Te. CUORE is also suitable to search for low energy rare events such as solar axions or WIMP scattering, thanks to its ultra-low background and large target mass. However, to conduct such sensitive searches requires improving the energy threshold to 10 keV. In this paper, we describe the analysis techniques developed for the low energy analysis of CUORE-like detectors, using the data acquired from No…
Next-generation experiments searching for neutrinoless double-beta decay must be sensitive to a Majorana neutrino mass as low as 10 meV. CUORE, an array of 988 TeO$_2$ bolometers being commissioned at Laboratori Nazionali del Gran Sasso in Italy, features an expected sensitivity of 50-130 meV at 90% C.L, that can be improved by removing the background from $α$ radioactivity. This is possible if, in coincidence with the heat release in a bolometer, the Cherenkov light emitted by the $β$ signal is detected. The amount of light detected is so far limited to only 100 eV, requiring low-noise cryoge…
The CUORE experiment will search for neutrinoless double-beta decay of $^{130}$Te with an array of 988 TeO$_2$ bolometers arranged in 19 towers. CUORE-0, the first tower assembled according to the CUORE procedures, was built and commissioned at Laboratori Nazionali del Gran Sasso, and took data from March 2013 to March 2015. In this paper we describe the design, construction and operation of the CUORE-0 experiment, with an emphasis on the improvements made over a predecessor experiment, Cuoricino. In particular, we demonstrate with CUORE-0 data that the design goals of CUORE are within reach.
Large mass bolometers are used in particle physics experiments to search for rare processes. By operating at low temperature, they are able to detect particle energies from few keV up to several MeV, measuring the temperature rise produced by the energy released. This study was performed on the bolometers of the CUORE experiment. The response function of these detectors is not linear in the energy range of interest, and it changes with the operating temperature. The non-linearity is found to be dominated by the thermistor and its biasing circuit. A method to obtain a linear response is the res…
Experiments searching for rare events, such as neutrinoless double beta decay and interactions of dark matter candidates, require extremely low levels of background. When these experiments are performed using macro-bolometers, radioactive contamination near the surfaces is of particular concern. In order to control the effects of surface contamination, we developed a novel technique that uses composite macro-bolometers to identify energy depositions that occur close to the surfaces of materials immediately surrounding the detector. The composite macro-bolometer proposed and studied here consis…
CUORE is a 741 kg array of 988 TeO$_2$ bolometeric crystals designed to search for the neutrinoless double beta decay of $^{130}$Te and other rare processes. CUORE has been taking data since summer 2017, and as of summer 2018 collected a total of 86.3 kg$\cdot$yr of TeO$_2$ exposure. Based on this exposure, we were able to set a limit on the $0νββ$ half-life of $^{130}$Te of $T^{0ν}_{1/2}>1.5\times10^{25}$ yr at 90% C.L. At this conference, we showed the decomposition of the CUORE background and were able to extract a $^{130}$Te $2νββ$ half-life of $T_{1/2}^{2ν}=[7.9\pm0.1 \mathrm{(stat.)}\pm0…
CUORE-0 is a cryogenic detector that uses an array of tellurium dioxide bolometers to search for neutrinoless double-beta decay of ^{130}Te. We present the first data analysis with 7.1 kg y of total TeO_2 exposure focusing on background measurements and energy resolution. The background rates in the neutrinoless double-beta decay region of interest (2.47 to 2.57 MeV) and in the α background-dominated region (2.70 to 3.90 MeV) have been measured to be 0.071 \pm 0.011 and 0.019 \pm 0.002 counts/keV/kg/y, respectively. The latter result represents a factor of 6 improvement from a predecessor expe…
We describe in detail the methods used to obtain the lower bound on the lifetime of neutrinoless double-beta ($0νββ$) decay in $^{130}$Te and the associated limit on the effective Majorana mass of the neutrino using the CUORE-0 detector. CUORE-0 is a bolometric detector array located at the Laboratori Nazionali del Gran Sasso that was designed to validate the background reduction techniques developed for CUORE, a next-generation experiment scheduled to come online in 2016. CUORE-0 is also a competitive $0νββ$ decay search in its own right and functions as a platform to further develop the anal…
The CUORE experiment, a ton-scale cryogenic bolometer array, recently began operation at the Laboratori Nazionali del Gran Sasso in Italy. The array represents a significant advancement in this technology, and in this work we apply it for the first time to a high-sensitivity search for a lepton-number--violating process: $^{130}$Te neutrinoless double-beta decay. Examining a total TeO$_2$ exposure of 86.3 kg$\cdot$yr, characterized by an effective energy resolution of (7.7 $\pm$ 0.5) keV FWHM and a background in the region of interest of (0.014 $\pm$ 0.002) counts/(keV$\cdot$kg$\cdot$yr), we f…
Coherent elastic neutrino-nucleus scattering (CEvNS) opens new approaches for the search of new physics beyond the Standard Model. The NUCLEUS experiment aims to use the intense antineutrino flux produced from nuclear reactor cores to perform measurements of the CEvNS cross-section via gram-scale ultra-low threshold cryogenic detectors. A common problem with low threshold detectors is the calibration, since most radioactive sources tend to saturate the dynamical range of the sensors. In this dissertation, a photon-statistic based optical calibration setup has been developed and tested using de…
Neutrinoless double beta decay (0nubb) is one of the most sensitive probes for physics beyond the Standard Model, providing unique information on the nature of neutrinos. In this paper we review the status and outlook for bolometric 0nubb decay searches. We summarize recent advances in background suppression demonstrated using bolometers with simultaneous readout of heat and light signals. We simulate several configurations of a future CUORE-like bolometer array which would utilize these improvements and present the sensitivity reach of a hypothetical next-generation bolometric 0nubb experimen…
CUPID-Mo is a bolometric experiment to search for neutrinoless double-beta decay ($0νββ$) of $^{100}$Mo. In this article, we detail the CUPID-Mo detector concept, assembly, installation in the underground laboratory in Modane in 2018, and provide results from the first datasets. The demonstrator consists of an array of 20 scintillating bolometers comprised of $^{100}$Mo-enriched 0.2 kg Li$_2$MoO$_4$ crystals. The detectors are complemented by 20 thin cryogenic Ge bolometers acting as light detectors to distinguish $α$ from $γ$/$β$ events by the detection of both heat and scintillation light si…
The CUPID-Mo experiment is searching for neutrinoless double beta decay in $^{100}$Mo, evaluating the technology of cryogenic scintillating Li$_{2}^{100}$MoO$_4$ detectors for CUPID (CUORE Upgrade with Particle ID). CUPID-Mo detectors feature background suppression using a dual-readout scheme with Li$_{2}$MoO$_4$ crystals complemented by Ge bolometers for light detection. The detection of both heat and scintillation light signals allows the efficient discrimination of $α$ from $γ$&$β$ events. In this proceedings, we discuss results from the first 2 months of data taking in spring 2019. In addi…