Development of a relaxation calorimeter for temperatures between 0.05 and 4 K

M. Brando

摘要:

本文详细介绍了一种适用于0.05 ~ 4K温度的等周量热计。所提出的设置可以提供小样品(通常为1 mg)的热容$C$的绝对值。基于蓝宝石衬底的样品平台的极其简单的设计以及仅使用锁定放大器和温度控制器的实验装置使得这种量热计的构造简单且廉价。采用热弛豫方法,该方法在样品平台和低温浴槽之间使用永久热连接$k$。$k(T)$的温度依赖性在整个温度范围内的几个平台上显示:$k(T)/T$几乎恒定到1 K,在那里它开始平稳下降。通过考虑平台组分的热阻,彻底解释了所观察到的行为。比较了基于蓝宝石和基于银的平台的$k(T)/T$值,未观察到显著差异。每个平台可以被组装成在1 K下具有特定的$k/T$值。由于样品的弛豫时间$τ\sim C/k$,$k(T)$可以调整到$C(T)$,以提供一个合理的快速测量时间:在这里,它演示了如何这种量热计可以用于所谓的单发制冷机($^{3}$他或退磁低温恒温器),其中一个单一的测量时间是有限的。此外,它可以用于中等磁场$B \leq 10$ T,因为平台成分是弱场依赖性的。

英文摘要:

A detailed description of an isoperibol calorimeter for temperatures between 0.05 and 4 K is presented. The proposed setup can provide absolute values of the heat capacity $C$ of small samples (typically 1 mg). The extremely simple design of the sample platform, based on a sapphire substrate, and the experimental setup, which makes use only of a lock-in amplifier and a temperature controller, make the construction of such a calorimeter easy and inexpensive. The thermal-relaxation method is employed, which utilizes a permanent thermal link $k$ between the sample platform and the low-temperature bath. The temperature dependence of $k(T)$ is shown for several platforms throughout the entire temperature range: $k(T)/T$ is nearly constant down to 1 K, where it starts to decrease smoothly. The observed behavior is thoroughly explained by considering the thermal resistances of the platform constituents. A comparison between the values of $k(T)/T$ for platforms based on sapphire and on silver is presented where no significant difference has been observed. Each platform can be assembled to have a particular value of $k/T$ at 1 K. Since the sample relaxation time $τ\sim C/k$, $k(T)$ can be adjusted to $C(T)$ to give a reasonably fast measuring time: Here, it is demonstrated how this calorimeter can be used in so-called single-shot refrigerators ($^{3}$He or demagnetization cryostats), where the time for a single measurement is limited. In addition, it can be used in moderate magnetic fields $B \leq 10$ T, because the platform constituents are weakly field dependent.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑