用于低温高压样品池的微型芯片量热计

Miniaturized chip calorimeter for high-pressure cells at low temperature

Neha Kondedan, Andreas Rydh

AI总结:

本文开发了一种微型纳米量热计,通过高频交流量热技术和组件堆叠实现微型化,适用于低温高压下的热容测量。

摘要:

高压下的热容测量对量热计提出了很高的要求。本文描述了一种用于低温高压热容测量的微型纳米量热计的开发。该器件在硅基底上制造,采用高频交流量热技术,外径为300 μm,厚度为25-40 μm,足够小以放入高压金刚石对顶砧中。通过将温度计和加热器等所有组件堆叠在中心区域内实现了微型化。薄膜量热计温度计尺寸为40 μm见方,保持了较大温度计的灵敏度和特性。制造过程采用可控各向异性刻蚀,制备出坚固性和厚度平衡的量热计芯片,适用于高压低温实验。量热计在10 Hz至1 kHz的较高特征频率下运行,将热振荡限制在以样品为主的有效体积内,从而避免了使用悬浮膜,这是传统纳米量热计的基础。

英文摘要:

Heat capacity measurements under high pressure places high demands on the calorimeter. Here we describe the development of a miniaturized nanocalorimeter for high-pressure heat capacity measurements at low temperature. The device, fabricated on a silicon substrate, employs a high-frequency AC calorimetry technique and features a design with an outer diameter of 300 $μ$m and thickness of 25-40 $μ$m, small enough to fit into high pressure diamond anvil cells. Miniaturization is achieved by stacking all components, including thermometer and heaters, within a central area. The thin-film calorimeter thermometer measures 40 $μ$m square and maintains the sensitivity and properties of larger thermometers. The fabrication process uses controlled anisotropic etch to produce calorimeter chips with a balance between robustness and thickness, suitable for experiments at high pressures and low temperatures. The calorimeter operates at a relatively high characteristic frequency between 10 Hz and 1 kHz, constraining the thermal oscillation to an effective volume dominated by the sample, thereby avoiding the use of a suspended membrane that is the basis for conventional nanocalorimeters.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑