We have successfully built and operated a source deployment system for the KamLAND detector. This system was used to position radioactive sources throughout the delicate 1-kton liquid scintillator volume, while meeting stringent material cleanliness, material compatibility, and safety requirements. The calibration data obtained with this device were used to fully characterize detector position and energy reconstruction biases. As a result, the uncertainty in the size of the detector fiducial volume was reduced by a factor of two. Prior to calibration with this system, the fiducial volume was the largest source of systematic uncertainty in measuring the number of anti-neutrinos detected by KamLAND. This paper describes the design, operation and performance of this unique calibration system.
Electrical capacitance volume sensor for microgravity mass gauging: Advancements in sensor calibration for microgravity fluid configurations and propellant management devices
Application of a high-precision distributed uranium source for determining the effective mass and volume of a HPGe detector
National Research Center «Kurchatov Institute»(库尔恰托夫国家研究中心);Joint Institute for Nuclear Research(苏联联合核研究所);Institute for Nuclear Research of the Russian Academy of Sciences(俄罗斯科学院核物理研究所);Kabardino-Balkarian State University named after H.M. Berbekov(H.M.别尔别科夫卡巴尔达-巴尔卡尔国立大学);Institute of Nuclear Physics of the Ministry of Energy of the Republic of Kazakhstan(哈萨克斯坦共和国能源部核物理研究所)