利用空间偏移拉曼光谱从离体人指骨预测腕部骨质疏松症:一项尸体研究
Predicting Wrist Osteoporosis from excised human finger bones using spatially offset Raman spectroscopy, A Cadaveric Study
AI总结:
本研究利用空间偏移拉曼光谱分析离体指骨,通过矿物质与基质比值差异,成功预测腕部DXA T评分(r=0.85),为骨质疏松症的无创诊断提供了新方法。
摘要:
骨质疏松症和骨量减少仍然被严重漏诊。目前的临床筛查几乎完全依赖双能X射线吸收测定法(DXA),该方法测量骨矿物质密度(BMD),但无法捕捉导致BMD流失的成分变化。我们研究了将空间偏移拉曼光谱(SORS)应用于离体指骨,能否评估能够诊断骨质疏松症和骨量减少并预测腕部DXA T评分的皮下生化标志物。在25具女性尸体的第二指近节指骨中段骨干上,以离体方式采集拉曼光谱,涵盖三个T评分类别(正常8例、骨量减少6例、骨质疏松11例),空间偏移分别为激光照射点的0、3和6毫米。将光谱归一化至PO43-峰后,在3毫米偏移处测量的三个类别的组平均光谱在CO32-、酰胺III、CH2和酰胺I谱带上显示出明显差异。定量上,五个矿物质与基质比值中的四个在正常骨与骨量减少骨之间以及骨量减少骨与骨质疏松骨之间均存在显著差异(p < 0.05),而所有五个比值在正常骨与骨质疏松骨之间均显示出显著差异。相比之下,0毫米偏移的对比度减弱,而6毫米偏移与3毫米偏移相比,并未增强不同组之间的区分度。基于3毫米光谱构建的留一法偏最小二乘回归模型预测桡骨远端DXA T评分的Pearson相关系数为r = 0.85,交叉验证均方根误差为1个T评分单位,正确分类了92%的标本。
英文摘要:
Osteoporosis and osteopenia remain vastly underdiagnosed. Current clinical screening relies almost exclusively on dual-energy X-ray absorptiometry (DXA), which measures bone mineral density (BMD) but fails to capture the compositional changes that lead to BMD loss. We investigated whether Spatially Offset Raman Spectroscopy (SORS) applied to excised finger bones can assess subsurface biochemical markers capable of diagnosing osteoporosis and osteopenia and predicting wrist DXA T-scores. Raman spectra were acquired ex vivo on the mid-shaft of the proximal phalanx of the second digit from 25 female cadavers spanning the three T-score categories (n=8 normal, n=6 osteopenic, and n=11 osteoporotic) at spatial offsets of 0, 3, and 6 mm from a laser irradiation spot. After normalizing spectra to the PO43- peak, group-averaged spectra of the three categories, measured at 3-mm offset, showed clear differences in the CO32-, Amide III, CH2, and Amide I bands. Quantitatively, four out of five mineral-to-matrix ratios differed significantly (p < 0.05) between normal and osteopenic bone, and between osteopenic and osteoporotic bone, and all five ratios showed significant differences between normal and osteoporotic bone. In contrast, the 0-mm offset suffered diminished contrast, and the 6-mm offset did not enhance discrimination between different groups, compared with the 3-mm offset. A leave-one-out, partial-least-squares regression model built from the 3-mm spectra predicted distal radius DXA T-score with a Pearson correlation of r = 0.85 and a root-mean-square error of cross-validation of 1 T-score units, correctly classifying 92% of specimens.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。