arXivDaily arXiv每日学术速递 周一至周五更新
arXiv周末暂无论文更新,休息一下吧,周末愉快~~
arXiv 2510.02847physics.med-ph
收藏

基于自旋锁磁共振成像的偶极序映射

Dipolar order mapping based on spin-lock magnetic resonance imaging

Department of Imaging and Interventional Radiology, Faculty of Medicine, The Chinese University of Hong Kong(香港中文大学医学院影像与介入放射学系);MR Research Collaboration, Siemens Heathineers Ltd.(西门子医疗磁共振研究合作部)

Zijian Gao, Qianxue Shan, Ziqin Zhou, Ziqiang Yu, Weitian Chen

AI总结:

本文提出一种基于自旋锁MRI技术的偶极序量化方法,通过引入特定比率RATIO_dosl,实现了对T1D和宏分子池分数的快速同步量化。

摘要:

目的:不均匀磁化转移(ihMT)效应反映偶极序与偶极弛豫时间(T1D),特指运动受限的大分子。我们旨在利用自旋锁MRI技术量化T1D。方法:在所提出的方法中,我们引入了一个特定于T1D的比率RATIO_dosl,该比率源自不同的弛豫率R_dosl,计算为双频弛豫R1ρ_dual与单频弛豫R1ρ_single测量的差值。开发了一种新型旋转回波自旋锁序列以实现双频自旋锁获取。建立了估计T1D以及宏分子池分数(MPF)图的框架。所提出的方法通过数值模拟、假体研究和在人体白质中的验证。结果:模拟显示RATIO_dosl对T1D的高灵敏度,并证实了所提出方法的准确性和鲁棒性。假体实验显示了鲁棒的ihMT对比度,并确认了通过RATIO_dosl量化T1D的能力。在体研究支持了该方法的临床可行性,仅使用三张自旋锁准备图像即可实现同时T1D和MPF映射。在十名健康志愿者中,白质T1D的平均值范围约为3.70至4.80毫秒。结论:本文提出了一种基于自旋锁MRI的T1D量化新方法。通过仅需三张对比度准备图像,该技术提供了一条有前景的路径,用于快速、稳健和同步的T1D和MPF量化,减少干扰因素。

英文摘要:

Purpose: Inhomogeneous magnetization transfer (ihMT) effect reflects dipolar order with a dipolar relaxation time ($T_{1D}$), specific to motion-restricted macromolecules. We aim to quantify $T_{1D}$ using spin-lock MRI technique. Methods: In the proposed method, we introduce a $T_{1D}$-specific ratio, denoted as $RATIO_{dosl}$. This ratio is derived from the distinct relaxation rate $R_{dosl}$, calculated as the difference between dual-frequency relaxation $R_{1ρ}^{dual}$ and single-frequency $R_{1ρ}^{single}$ relaxation measurements. A novel rotary-echo spin-lock sequence was developed to enable dual-frequency spin-lock acquisition. We established a framework to estimate $T_{1D}$, as well as the macromolecular pool fraction (MPF) map. The proposed approach was validated via numerical simulations, phantom studies, and demonstrated in vivo in human white matter. Results: Simulations revealed the high sensitivity of $RATIO_{dosl}$ to $T_{1D}$, and substantiated the accuracy and robustness of the proposed methods. Phantom experiments demonstrated robust ihMT contrast and confirmed the capability of $T_{1D}$ quantification via $RATIO_{dosl}$. In vivo studies supported the clinical viability of this approcah, achieving simultaneous $T_{1D}$ and MPF mapping using only three spin-lock prepared images. Across ten healthy volunteers, the mean white matter $T_{1D}$ ranged from approximately 3.70 to 4.80 ms. Conclustion: We propose a novel method for $T_{1D}$ quantification based on spin-lock MRI. By requiring only three contrast-prepared images, this technique provides a promising pathway for robust, rapid, and simultaneous $T_{1D}$ and MPF quantification with fewer confounds

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑