arXiv 2408.09733physics.med-ph
收藏

通过抑制残余偶极耦合实现组织中大分子质子分数的方向无关定量测量

Orientation independent quantification of macromolecular proton fraction in tissues with suppression of residual dipolar coupling

The Chinese University of Hong Kong(香港中文大学);Siemens Heathineers Ltd.(西门子医疗有限公司);Illumination Medical Technology Limited(亮明医疗科技有限公司)

Zijian Gao, Ziqiang Yu, Ziqin Zhou, Jian Hou, Baiyan Jiang, Michael Ong, Weitian Chen

AI总结:

针对有序组织中定量MT成像的RDC效应导致的方向依赖性问题,本研究提出MPF-SL技术,通过设计偏共振自旋锁定脉冲抑制RDC,经标本及人体实验验证其方向敏感性更低,可实现方向无关的MPF定量测量。

摘要:

定量磁化传递(MT)成像可对组织的大分子环境进行无创表征。但近期研究表明,有序组织结构中MT参数的定量测量存在方向依赖性,可能会对其临床应用造成干扰。值得注意的是,在关节软骨、髓鞘等具有有序结构的组织中,水质子的偶极-偶极相互作用未被完全平均,会产生残余偶极耦合(RDC)效应。本研究证明,采用一种名为基于自旋锁定的大分子质子分数成像(MPF-SL,spin-lock 即自旋锁定)的新兴技术,可抑制RDC对有序组织定量MT成像的干扰效应。MPF-SL中的偏共振自旋锁定脉冲可被设计为产生强有效自旋锁定场,在不违反临床扫描的比吸收率及硬件限制的前提下抑制RDC。此外,MPF-SL中去除水信号的设计,使得强有效自旋锁定场的应用不会受到直接水饱和带来的干扰信号影响。我们的研究结果通过人类膝关节标本及健康人体软骨实验得到了验证。结果表明,与R2、R1rho及基于饱和脉冲的MT成像相比,MPF-SL对组织方向的敏感性更低。因此,MPF-SL可作为一种有价值的方向无关技术用于MPF定量测量。

英文摘要:

Quantitative magnetization transfer (MT) imaging enables non-invasive characterization of the macromolecular environment of tissues. However, recent work has highlighted that the quantification of MT parameters exhibits orientation dependence in ordered tissue structures, potentially confounding its clinical applications. Notably, in tissues with ordered structures, such as articular cartilage and myelin, the residual dipolar coupling (RDC) effect can arise owing to incomplete averaging of dipolar-dipolar interactions of water protons. In this study, we demonstrated the confounding effect of RDC on quantitative MT imaging in ordered tissues can be suppressed by using an emerging technique known as macromolecular proton fraction mapping based on spin-lock (MPF-SL). The off-resonance spin-lock pulse in MPF-SL could be designed to generate a strong effective spin-lock field to suppress RDC without violating the specific absorption rate and hardware limitations in clinical scans. Furthermore, removing the water signal in MPF-SL enabled the application of a strong effective spin-lock field without any confounding signal from direct water saturation. Our findings were experimentally validated using human knee specimens and healthy human cartilage. The results demonstrated that MPF-SL exhibits lower sensitivity to tissue orientation compared with R2, R1rho, and saturation-pulse-based MT imaging. Thus, MPF-SL could serve as a valuable orientation-independent technique for quantifying MPF.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑