arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 2409.20194physics.ins-det
收藏

键合至 Timepix3 读出芯片的 55 μm 节距 iLGAD 传感器的实验室与束流测试性能研究

Laboratory and beam-test performance study of a 55$~μ$m pitch iLGAD sensor bonded to a Timepix3 readout chip

Peter Svihra, Richard Bates, Justus Braach, Eric Buschmann, Dominik Dannheim, Dima Maneuski, Neil Moffat, Younes Otarid

AI总结:

本文表征键合 Timepix3 的 55 μm 节距、2 cm² iLGAD,通过 X 射线荧光校准和 CERN 120 GeV/c π 介子束测试,验证其均匀增益、时间与能量分辨率。

摘要:

本投稿报告了一款大面积(2 cm²)小节距(55 μm)反向低增益雪崩探测器(inverse Low-Gain Avalanche Detector,iLGAD)的表征结果,该传感器键合至一块 Timepix3 读出芯片。iLGAD 传感器由 Micron Semiconductor Ltd 生产,目标是获得良好的增益均匀性并最大化填充因子——这是标准小节距 LGAD 设计中存在的问题。我们利用 X 射线校准和束流测试开展了详细的性能评估。X 射线荧光装置被用于获得能量校准并确定新器件的最佳工作设置,而广泛的束流测试则实现了对探测器性能的详细评估。束流测试在 CERN SPS 北区 H6 束流线进行,使用 120 GeV/c 的 π 介子束。参考径迹测量和时间戳由基于 Timepix3 的束流望远镜装置实现。结果表明,在整个增益区域测得约 5 的增益且均匀性非常好,同时在未校正时间游走效应的情况下,击中时间分辨率可达 1.3 ns。此外,结果表明该增益使低至 4.5 keV 的良好 X 射线能量分辨率成为可能。

英文摘要:

This contribution reports on characterisation results of a large-area (2$~\mathrm{cm}^2$) small pitch (55$~μ$m) inverse Low-Gain Avalanche Detector (iLGAD), bonded to a Timepix3 readout chip. The ilGAD sensors were produced by Micron Semiconductor Ltd with the goal to obtain good gain uniformity and maximise the fill-factor -- an issue present with standard small-pitch LGAD designs. We have conducted detailed performance evaluations using both X-ray calibrations and beam tests. An X-ray fluorescence setup has been used to obtain energy calibration and to identify the optimal operating settings of the new devices, whereas the extensive beam tests allowed for a detailed evaluation of the detector performance. The beam-tests were performed at the CERN SPS North Area H6 beamline, using a 120 GeV/c pion beam. The reference tracking and time-stamping is achieved by a Timepix3-based beam telescope setup. The results show a gain of around 5 with very good uniformity, measured across the whole gain area, as well as a hit time resolution down to 1.3 ns without correcting for the time-walk effects. Furthermore, it is shown that the gain opens the possibility of a good X-ray energy resolution down to 4.5 keV.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑