基于部分穆勒偏振测量数据的诊断偏振图像完整重建的机器学习模型

Machine Learning Model for Complete Reconstruction of Diagnostic Polarimetric Images from partial Mueller polarimetry data

Sooyong Chae, Tongyu Huang, Omar Rodrıguez-Nunez, Théotim Lucas, Jean-Charles Vanel, Jérémy Vizet, Angelo Pierangelo, Gennadii Piavchenko, Tsanislava Genova, Aj… 展开作者

Sooyong Chae, Tongyu Huang, Omar Rodrıguez-Nunez, Théotim Lucas, Jean-Charles Vanel, Jérémy Vizet, Angelo Pierangelo, Gennadii Piavchenko, Tsanislava Genova, Ajmal Ajmal, Jessica C. Ramella-Roman, Alexander Doronin, Hui Ma, Tatiana Novikova

AI总结:

针对部分穆勒偏振测量无法获取完整矩阵的问题,提出基于序贯神经网络的机器学习模型,从测得的前三行重建缺失元素,实现视频速率下的高精度偏振图像重建,用于临床医学诊断。

摘要:

将成像穆勒偏振测量法转化为临床实践,常常受到现有仪器体积大、采集速度相对较慢的阻碍。使用偏振敏感相机作为探测器可以减小仪器尺寸,并允许以视频速率进行数据流传输。然而,只能测量完整的4x4穆勒矩阵的前三行。为了克服这一障碍,我们开发了一种机器学习方法,使用序贯神经网络算法,从测得的前三行元素中重建穆勒矩阵的缺失元素。该算法在两种不同成像穆勒偏振计采集的各种离体人体组织(子宫颈、结肠、皮肤、脑)的偏振图像数据集上进行了训练和测试,这两种偏振计分别工作在反射(宽场成像系统)或透射(显微镜)配置下,波长分别为550纳米和385纳米。使用各种误差指标评估重建性能,所有指标均确认误差值较低。训练好的神经网络算法对单个图像像素的执行时间约为300微秒。这表明,将并行处理所有图像像素的机器学习方法与以视频速率运行的部分穆勒偏振计相结合,可以有效替代完整的穆勒偏振计,并生成生物组织的去偏振、线性延迟和光轴方向的精确图谱,可用于临床环境中的医学诊断。

英文摘要:

The translation of imaging Mueller polarimetry to clinical practice is often hindered by large footprint and relatively slow acquisition speed of the existing instruments. Using polarization-sensitive camera as a detector may reduce instrument dimensions and allow data streaming at video rate. However, only the first three rows of a complete 4x4 Mueller matrix can be measured. To overcome this hurdle we developed a machine learning approach using sequential neural network algorithm for the reconstruction of missing elements of a Mueller matrix from the measured elements of the first three rows. The algorithm was trained and tested on the dataset of polarimetric images of various excised human tissues (uterine cervix, colon, skin, brain) acquired with two different imaging Mueller polarimeters operating in either reflection (wide-field imaging system) or transmission (microscope) configurations at different wavelengths of 550 nm and 385 nm, respectively. The reconstruction performance was evaluated using various error metrics, all of which confirmed low error values. The execution time of the trained neural network algorithm was about 300 microseconds for a single image pixel. It suggests that a machine learning approach with parallel processing of all image pixels combined with the partial Mueller polarimeter operating at video rate can effectively substitute for the complete Mueller polarimeter and produce accurate maps of depolarization, linear retardance and orientation of the optical axis of biological tissues, which can be used for medical diagnosis in clinical settings.