arXiv 2207.01481physics.opticseess.SP
收藏

消除偏振误差后高精度(<0.004 dB)测量的光连接器的偏振相关损耗

Polarization-Dependent Loss of Optical Connectors Measured with High Accuracy (<0.004 dB) after Cancelation of Polarimetric Errors

Reinhold Noe, Benjamin Koch

摘要:

国家的最先进的旋光仪校准进行了审查。在不改变功率的情况下产生许多准随机偏振状态并移动/弯曲光纤允许找到偏振度达到统一并且寄生偏振相关损耗小的偏振计校准。使用偏振扰频器/Transformer和偏振计,可以表征被测器件。其Mueller矩阵可以分解为非去极化Mueller-Jones矩阵乘以纯去极化Mueller矩阵的乘积。测试极化可能随时间漂移。在光开关的帮助下,参考器件可以根据内部参考路径进行测量。之后,在可能存在不同测试偏振的情况下,实际被测器件可以根据内部参考进行测量。因此克服了偏振漂移和重复参考器件测量的需要。当插入插接线时,可以测量连接器PDL,前提是再次通过光纤移动/弯曲来校准误差。在实验上,我们测量了PDL的误差<0.004 dB。这很容易足以测量连接器PDL,这是证明。当被测器件是良好的偏振器时,测量到PDL >60 dB。一个20 Mrad/s的偏振扰码器与铌酸锂器件产生的测试偏振。该偏振计可以在100 MHz的频率下进行采样,并可以存储64 M斯托克斯矢量。在激光频率扫描期间,可以在短至5 μ s的时间间隔内测量Mueller矩阵。

英文摘要:

State-of-the-art polarimeter calibration is reviewed. Producing many quasi-random polarization states and moving/bending a fiber without changing power allows finding a polarimeter calibration where the degree-of-polarization reaches unity and parasitic polarization-dependent loss is small. Using a polarization scrambler/transformer and a polarimeter a device-under-test can be characterized. Its Mueller matrix can be decomposed into a product of a nondepolarizing Mueller-Jones matrix times a purely depolarizing Mueller matrix. Test polarizations may drift over time. With help of an optical switch the reference device can be measured against an internal reference path. Later, with possibly different test polarizations, the actual device-under-test is measured against the internal reference. Polarization drift and need for repeated reference device measurement are thus overcome. When a patchcord is inserted, connector PDL can be measured, provided that errors are calibrated away, again by fiber moving/bending. Experimentally we have measured PDL with errors <0.004 dB. This easily suffices to measure connector PDL, which is demonstrated. PDL >60 dB was measured when the device under test was a good polarizer. A 20 Mrad/s polarization scrambler with LiNbO3 device generates the test polarizations. The polarimeter can sample at 100 MHz and can store 64M Stokes vectors. During laser frequency scans Mueller matrices can be measured in time intervals as short as 5 us.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑