深部脑超声消融热剂量建模及体内实验验证

Deep Brain Ultrasound Ablation Thermal Dose Modeling with in Vivo Experimental Validation

Worcester Polytechnic Institute(伍斯特理工学院);Albany Medical Center(奥尔巴尼医学中心);GE Global Research Center(通用电气全球研究中心);Charles E. Schmidt College of Medicine, Florida Atlantic University(佛罗里达大西洋大学查尔斯·E·施密特医学院);Acoustic MedSystems, Inc.(声学医疗系统公司)
Zhanyue Zhao, Benjamin Szewczyk, Matthew Tarasek, Charles Bales, Yang Wang, Ming Liu, Yiwei Jiang, Chitresh Bhushan, Eric Fiveland, Zahabiya Campwala, Rachel Tr… 展开作者

Zhanyue Zhao, Benjamin Szewczyk, Matthew Tarasek, Charles Bales, Yang Wang, Ming Liu, Yiwei Jiang, Chitresh Bhushan, Eric Fiveland, Zahabiya Campwala, Rachel Trowbridge, Phillip M. Johansen, Zachary Olmsted, Goutam Ghoshal, Tamas Heffter, Katie Gandomi, Farid Tavakkolmoghaddam, Christopher Nycz, Erin Jeannotte, Shweta Mane, Julia Nalwalk, E. Clif Burdette, Jiang Qian, Desmond Yeo, Julie Pilitsis, Gregory S. Fischer

AI总结:

本文针对脑部NBTU热消融,构建融合声压场、生物热传递与损伤评估的FEM模型,并经体内MRTI验证,温度和消融体积误差及相关性指标均显示模拟准确。

摘要:

体内基于针头的治疗性超声(NBTU)是介入恶性脑肿瘤的一种微创选择,常用于热消融手术。该技术适用于原发性和转移性癌症,利用最高达10 MHz的高频交变电场激励压电换能器。换能器随之产生的快速形变形成声波,在组织中传播,使靶肿瘤部位出现局部高温加热并诱导细胞快速死亡。为优化NBTU换能器在治疗期间递送热剂量的设计,研究者经常对形变压电换能器产生的声压场进行数值建模。由输入压力场产生的生物热传递过程用于追踪施加器随时间的热传播。磁共振热成像(MRTI)可用于对这些模型进行实验验证。使用MRTI得到的验证结果证明了该模型的可行性,并显示出一致的热传播模式。然而,热损伤等剂量图更有利于评估治疗效果。为了在真实脑组织环境基础上获得更准确的模拟,本研究开展了一项具有增强损伤评估能力的新型有限元法(FEM)模拟。结果显示,实验与模拟结果之间的最高温度差异为2.1884°C(3.71%),消融体积差异为0.0631 cm³(5.74%)。峰值温度的最低Pearson相关系数(PCC)为0.7117,消融区域的最低Dice系数为0.7021,表明模拟与实验在准确性上具有良好一致性。

英文摘要:

Intracorporeal needle-based therapeutic ultrasound (NBTU) is a minimally invasive option for intervening in malignant brain tumors, commonly used in thermal ablation procedures. This technique is suitable for both primary and metastatic cancers, utilizing a high-frequency alternating electric field (up to 10 MHz) to excite a piezoelectric transducer. The resulting rapid deformation of the transducer produces an acoustic wave that propagates through tissue, leading to localized high-temperature heating at the target tumor site and inducing rapid cell death. To optimize the design of NBTU transducers for thermal dose delivery during treatment, numerical modeling of the acoustic pressure field generated by the deforming piezoelectric transducer is frequently employed. The bioheat transfer process generated by the input pressure field is used to track the thermal propagation of the applicator over time. Magnetic resonance thermal imaging (MRTI) can be used to experimentally validate these models. Validation results using MRTI demonstrated the feasibility of this model, showing a consistent thermal propagation pattern. However, a thermal damage isodose map is more advantageous for evaluating therapeutic efficacy. To achieve a more accurate simulation based on the actual brain tissue environment, a new finite element method (FEM) simulation with enhanced damage evaluation capabilities was conducted. The results showed that the highest temperature and ablated volume differed between experimental and simulation results by 2.1884°C (3.71%) and 0.0631 cm$^3$ (5.74%), respectively. The lowest Pearson correlation coefficient (PCC) for peak temperature was 0.7117, and the lowest Dice coefficient for the ablated area was 0.7021, indicating a good agreement in accuracy between simulation and experiment.