利用DEPFET进行低温质子辐照以研究Athena宽视场成像仪

Low temperature proton irradiation with DEPFETs for Athena's Wide Field Imager

Valentin Emberger, Robert Andritschke, Parviz Azhdarzadeh, Günter Hauser, Astrid Mayr, Johannes Müller-Seidlitz, Abbas Rezaei, Wolfgang Treberer-Treberspurg

AI总结:

本文通过用62.4 MeV质子辐照Athena WFI的DEPFET原型探测器,研究位移损伤导致的暗电流增加及低温退火效应,为评估其在空间辐射环境下的性能退化提供关键数据。

摘要:

宽视场成像仪(WFI)是欧空局下一代大型X射线任务Athena上的两台仪器之一,旨在以宽视场和高计数率能力对0.2至15 keV范围内的X射线进行成像光谱观测。其焦平面由背照式DEPFET(耗尽型p沟道场效应晶体管)传感器组成,这些传感器具有高辐射耐受性,并能提供接近Fano极限的能量分辨率。为实现这一目标,需要极低噪声的读出,预计在寿命初期约为3个电子ENC。这使得该器件对任何由辐射引起的读出噪声恶化都非常敏感。性能退化的主要机制将是暗电流的增加,而暗电流增加主要由高能质子造成的位移损伤引起。为研究预期的性能退化,一个具有完全代表性像素布局和制造技术的原型探测器模块在维也纳新城MedAustron加速器设施中接受了62.4 MeV质子的辐照。总剂量相当于3.3×10^9个10-MeV质子/cm²,分两步施加。在辐照期间、辐照间隔以及辐照之后,探测器始终保持在213 K的工作温度下,并保持完全偏置和运行。记录了数据以分析所有入射粒子的信号。我们报告了辐照后暗电流的增加情况,并给出了213 K下与电流相关的损伤率。此外,还展示了在213 K、236 K、253 K、273 K和289 K下低温退火的效果。

英文摘要:

The Wide Field Imager (WFI), one of two instruments on ESA's next large X-ray mission Athena, is designed for imaging spectroscopy of X-rays in the range of 0.2 to 15 keV with a large field of view and high count rate capability. The focal plane consists of back-illuminated DEPFET (Depleted p-channel field effect transistor) sensors that have a high radiation tolerance and provide a near Fano-limited energy resolution. To achieve this, a very low noise readout is required, about 3 electrons ENC at beginning of life is foreseen. This makes the device very susceptible to any radiation induced worsening of the readout noise. The main mechanism of degradation will be the increase of dark current due to displacement damage caused primarily by high energy protons. To study the expected performance degradation, a prototype detector module with fully representative pixel layout and fabrication technology was irradiated with 62.4 MeV protons at the accelerator facility MedAustron in Wiener Neustadt. A total dose equivalent to 3.3 $\textstyle{10^{9}}$ 10-MeV protons/$\mathrm{cm^{2}}$ was applied in two steps. During, in-between and after the irradiations the detector remained at the operating temperature of 213 K and was fully biased and operated. Data was recorded to analyze the signal of all incident particles. We report on the increase of dark current after the irradiation and present the current related damage rate at 213 K. The effect of low temperature annealing at 213 K , 236 K, 253 K, 273 K, and 289 K is presented.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑