超快电子显微镜中的Wehnelt光电发射:稳定性与可用性
Wehnelt Photoemission in an Ultrafast Electron Microscope: Stability and Usability
AI总结:
本研究测试比较超快电子显微镜中三种阴极构型的稳定性与可用性,发现Wehnelt电极Ni光阑离轴光电发射的稳定性和束流表现最优,为UEM实验及模式切换提供参考。
摘要:
我们在基于热电子发射的超快电子显微镜(UEM)中,测试并比较了三种不同阴极材料与构型的稳定性和可用性:(1)带石墨保护环、直径0.1 mm的LaB6源的轴上热电子发射与光电发射;(2)Wehnelt电极的Ni光阑表面的离轴光电发射;(3)直径0.2 mm的多晶Ta源的轴上热电子发射与光电发射。针对每种阴极类型与构型,我们阐明了热电子加热后源的同步冷却会对光电子束流稳定性产生有害影响。此外,我们通过采集平行束与会聚束电子衍射图样、以及形成最优探针尺寸,验证了其可用性。我们发现,离轴Ni Wehnelt光阑光电发射的可用性至少与轴上LaB6热电子发射、以及轴上光电发射相当。然而,无论电子发射机制如何,Wehnelt光阑离轴光电发射的稳定性和可达到的束流均优于其他阴极类型与构型。该构型的束流稳定性在70分钟(测试的最长时长)内保持在平均值的1%以内;对于15 pA的束流(即200 kHz重复频率下每束团460个电子),稳态束流在本文采用的采样时间分辨率(~1 s)内即可达到。稳态条件的重复性和鲁棒性也保持在平均值的1%以内。我们讨论了这些发现对UEM成像与衍射实验、脉冲束损伤测量、以及在同一仪器中最优常规TEM与UEM运行模式间实际切换的意义。
英文摘要:
We tested and compared the stability and usability of three different cathode materials and configurations in a thermionic-based ultrafast electron microscope: (1) on-axis thermionic and photoemission from a 0.1-mm diameter LaB6 source with graphite guard ring, (2) off-axis photoemission from the Ni aperture surface of the Wehnelt electrode, and (3) on-axis thermionic and photoemission from a 0.2-mm diameter polycrystalline Ta source. For each cathode type and configuration, we illustrate how the photoelectron beam-current stability is deleteriously impacted by simultaneous cooling of the source following thermionic heating. Further, we demonstrate usability via collection of parallel- and convergent-beam electron diffraction patterns and by formation of optimum probe size. We find that usability of the off-axis Ni Wehnelt-aperture photoemission is at least comparable to on-axis LaB6 thermionic emission, as well as to on-axis photoemission. However, the stability and achievable beam currents for off-axis photoemission from the Wehnelt aperture were superior to that of the other cathode types and configurations, regardless of the electron-emission mechanism. Beam-current stability for this configuration was found to be within 1% of the mean for 70 minutes (longest duration tested), and steady-state beam current was reached within the sampling-time resolution used here (~1 s) for 15 pA beam currents (i.e., 460 electrons per packet for a 200 kHz repetition rate). Repeatability and robustness of the steady-state condition was also found to be within 1% of the mean. We discuss the implications of these findings for UEM imaging and diffraction experiments, for pulsed-beam damage measurements, and for practical switching between optimum conventional TEM and UEM operation within the same instrument.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。