超声颈脊髓神经调控的策略与安全性仿真
Strategies and safety simulations for ultrasonic cervical spinal cord neuromodulation
AI总结:
本文通过声热联合建模仿真500 kHz聚焦超声经后部声窗和外侧椎间孔作用于颈椎C5/C6的神经调控,揭示个体间压力与升温差异显著,强调需开发跨脊柱超声仿真软件以保障治疗安全有效。
摘要:
聚焦超声脊髓神经调控研究已在小动物中证明了脊髓神经调控效果。要将这些方法安全有效地转化到人体尺度,需要理解超声波在人体脊柱内的传播和热量沉积。为此,本研究采用声学与热学联合建模,评估一个500 kHz声源聚焦到颈椎C5/C6节段时产生的压力和热分布,并通过两种路径实现:a)椎后弓之间的后部声窗,或b)C6脊神经穿出的外侧椎间孔。针对四名受试者的CT体积,以及±10 mm平移和±10°旋转的声源定位误差,仿真了150个0.1 s脉冲组成的脉冲串,脉冲重复频率为0.33 Hz,自由场空间峰值脉冲平均强度为10 W/cm^2。采用后部路径时,目标压力为自由场空间峰值压力的20%至70%;采用外侧路径时为20%至100%。当声源以后部路径最优定位时,脊柱峰值升温低于1°C,但声源定位偏差会导致骨骼升温高达4°C。在定位偏差范围内,外侧路径的升温不超过2°C。不同受试者之间的目标压力和峰值升温存在显著差异。根据路径不同,受试者之间的目标压力变化达三到四倍,而峰值升温在受试者之间变化约两倍。这导致受试者之间每摄氏度峰值升温所获得的目标压力相差近十倍。这凸显了开发跨脊柱超声仿真软件的重要性,以确保聚焦超声脊髓疗法的受试者特异性安全性和有效性。
英文摘要:
Focused ultrasound spinal cord neuromodulation studies have demonstrated spinal cord neuromodulation in small animals. The safe and efficacious translation of these approaches to human scale requires an understanding of ultrasound propagation and heat deposition within the human spine. To address this, combined acoustic and thermal modelling was used to assess the pressure and heat distributions produced by a 500 kHz source focused to the C5/C6 level of the cervical spine via two approaches a) the posterior acoustic window between vertebral posterior arches, or b) the lateral intervertebral foramen from which the C6 spinal nerve exits. Pulse trains of 150 0.1 s pulses with a pulse repetition frequency of 0.33 Hz and free-field spatial peak pulse-averaged intensity of 10 W/cm^2 were simulated for the CT volumes of four subjects and for $\pm$10 mm translational and $\pm$10° rotational source positioning errors. Target pressures ranged between 20% and 70% of free-field spatial peak pressures with the posterior approach, and 20% and 100% with the lateral approach. When the source was optimally positioned with the posterior approach, peak spine heating values were below 1°C, but source mis-positioning resulted in bone heating up to 4°C. Heating with the lateral approach did not exceed 2°C within the mispositioning range. There were substantial inter-subject differences in target pressures and peak heating values. Target pressure varied three to four-fold between subjects, depending on approach, while peak heating varied approximately two-fold between subjects. This results in a near ten-fold range in the target pressure achieved per degree of peak heating between subjects. This highlights the importance of developing trans-spine ultrasound simulation software for the assurance of subject-specific safety and efficacy of focused ultrasound spinal cord therapies.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。