AI总结:
针对动态光纤弯曲导致的波前畸变影响多芯光纤无透镜成像的问题,提出基于仅强度测量的叠层成像无透镜相干内窥显微术,以扫描纤芯照明时的反射衍射强度图为输入恢复物体相位与振幅,采集流程等效于共焦显微内窥镜且仅替换探测器为相机。
摘要:
传统基于光纤束的内窥镜通过在远端放置微型透镜,并将每根纤芯用作成像像素,可借助柔性多芯光纤(MCF)束实现微创成像。近年来,通过校正校准流程中预先测量的纤芯间相位畸变,经多芯光纤(MCF)的无透镜成像成为可能。然而,随时间变化的波前畸变(例如由动态光纤弯曲引起的畸变)对此类方法构成了挑战。本文中,我们展示了一种基于仅强度测量的相干无透镜成像技术,该技术对纤芯间相位畸变不敏感。我们利用叠层成像重建算法,以照明光在多芯光纤(MCF)纤芯上扫描时从物体反射的一组衍射强度图案作为输入,来恢复放置在距光纤尖端一定距离处的反射式物体的相位和振幅分布。因此,我们的方法采用与共焦显微内窥镜等效的采集流程,仅将单探测器替换为相机。
英文摘要:
Conventional fiber-bundle-based endoscopes allow minimally invasive imaging through flexible multi-core fiber (MCF) bundles by placing a miniature lens at the distal tip and using each core as an imaging pixel. In recent years, lensless imaging through MCFs was made possible by correcting the core-to-core phase distortions pre-measured in a calibration procedure. However, temporally varying wavefront distortions, for instance, due to dynamic fiber bending, pose a challenge for such approaches. Here, we demonstrate a coherent lensless imaging technique based on intensity-only measurements insensitive to core-to-core phase distortions. We leverage a ptychographic reconstruction algorithm to retrieve the phase and amplitude profiles of reflective objects placed at a distance from the fiber tip, using as input a set of diffracted intensity patterns reflected from the object when the illumination is scanned over the MCF cores. Our approach thus utilizes an acquisition process equivalent to confocal microendoscopy, only replacing the single detector with a camera.