arXiv 2311.05412physics.med-ph
收藏

使用定量指标和机器学习预测放射科医生对MRI图像质量及伪影的解读

The Use of Quantitative Metrics and Machine Learning to Predict Radiologist Interpretations of MRI Image Quality and Artifacts

Lucas McCullum, John Wood, Maria Gule-Monroe, Ho-Ling Anthony Liu, Melissa Chen, Komal Shah, Noah Nathan Chasen, Vinodh Kumar, Ping Hou, Jason Stafford, Carolin… 展开作者

Lucas McCullum, John Wood, Maria Gule-Monroe, Ho-Ling Anthony Liu, Melissa Chen, Komal Shah, Noah Nathan Chasen, Vinodh Kumar, Ping Hou, Jason Stafford, Caroline Chung, Moiz Ahmad, Christopher Walker, Joshua Yung

AI总结:

本研究利用MRQy工具提取MRI定量指标,结合放射科医生评估训练bagging集成决策树模型,预测图像质量与伪影,发现通用模型表现略优于单一模型,有望用于临床实时图像质量自动检查。

摘要:

作为质量改进项目的一部分,收集了3D-GRE和3D-TSE脑部3T对比后T1加权图像数据集,并将其展示给五名神经放射科医生,由他们对每个序列的图像质量和成像伪影进行评估。使用MRQy工具对相同的扫描进行处理,以获取客观、定量的图像质量指标。利用结合放射科医生评估和定量指标的数据集,训练了一个采用bagging集成方法的决策树分类器,以使用定量指标预测放射科医生的评估结果。针对以下三个任务开发了机器学习模型:(1)确定每种MRI序列和评估指标的最佳模型/性能,(2)确定所有MRI序列中针对每种评估指标的最佳模型/性能,(3)确定所有MRI序列和评估中通用的最佳模型/性能。成像伪影的模型性能略高于图像质量,例如,图像质量的最终通用模型AUROC为0.77(0.41 - 0.84,95% CI),而成像伪影为0.78(0.60 - 0.93,95% CI)。此外,注意到通用模型的性能略优于单个模型(例如,3D-GRE图像质量的AUROC为0.69),表明了综合训练数据在这些应用中的价值。这些模型可部署在诊所中,作为实时图像采集的自动检查,以防止在回顾性放射科医生分析后需要重新安排预约进行患者重新扫描,或提高阅片者对研究的信心。需要进一步在外部数据集上验证此处描述的模型。此处展示的结果表明,MRQy可作为定量指标的基础,作为放射科医生评估的替代方案。

英文摘要:

A dataset of 3D-GRE and 3D-TSE brain 3T post contrast T1-weighted images as part of a quality improvement project were collected and shown to five neuro-radiologists who evaluated each sequence for both image quality and imaging artifacts. The same scans were processed using the MRQy tool for objective, quantitative image quality metrics. Using the combined radiologist and quantitative metrics dataset, a decision tree classifier with a bagging ensemble approach was trained to predict radiologist assessment using the quantitative metrics. A machine learning model was developed for the following three tasks: (1) determine the best model / performance for each MRI sequence and evaluation metric, (2) determine the best model / performance across all MRI sequences for each evaluation metric, and (3) determine the best general model / performance across all MRI sequences and evaluations. Model performance for imaging artifact was slightly higher than image quality, for example, the final generalized model AUROC for image quality was 0.77 (0.41 - 0.84, 95% CI) while imaging artifact was 0.78 (0.60 - 0.93, 95% CI). Further, it was noted that the generalized model performed slightly better than the individual models (AUROC 0.69 for 3D-GRE image quality, for example), indicating the value in comprehensive training data for these applications. These models could be deployed in the clinic as automatic checks for real-time image acquisition to prevent patient re-scanning requiring another appointment after retrospective radiologist analysis or improve reader confidence in the study. Further work needs to be done to validate the model described here on an external dataset. The results presented here suggest that MRQy could be used as a foundation for quantitative metrics as a surrogate for radiologist assessment.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑